小波域并行MRI SENSE重建的迭代法
An Iterative Method for Parallel MRI SENSE-based Reconstruction in the Wavelet Domain
摘要:
为了在一些MRI应用中减少扫描时间和/或提高空间/时间分辨率,自20世纪90年代初以来,具有多线圈采集的并行MRI(pMRI)采集技术已经作为允许更快地采集减小的视场(FOV)图像的强大3D成像方法出现。在这些技术中,必须从所得到的采集的欠采样k空间数据重建完整FOV图像。为此,已经提出了几种重建技术,例如广泛使用的SENSE方法。然而,当在测量数据和估计线圈灵敏度图两者中发生扰动时,重建图像通常呈现伪影。在本文中,我们的目标是实现良好的重建图像质量时,使用低磁场和高缩减因子。在这些严格的实验条件下,SENSE方法和图像域中的Tikhonov正则化都不能给出令人信服的结果。为此,我们提出了一种新的方法,在复小波域的正则化进行基于SENSE的重建。为了进一步提高重建图像的质量,在正则化过程中加入局部凸约束。在1.5特斯拉的磁共振回波(GRE)解剖和回波平面成像(EPI)功能磁共振成像数据进行的体内实验表明,该算法提供了重建图像与减少的伪影高缩减因子。
英文摘要:
To reduce scanning time and/or improve spatial/temporal resolution in some MRI applications, parallel MRI (pMRI) acquisition techniques with multiple coils acquisition have emerged since the early 1990s as powerful 3D imaging methods that allow faster acquisition of reduced Field of View (FOV) images. In these techniques, the full FOV image has to be reconstructed from the resulting acquired undersampled k-space data. To this end, several reconstruction techniques have been proposed such as the widely-used SENSE method. However, the reconstructed image generally presents artifacts when perturbations occur in both the measured data and the estimated coil sensitivity maps. In this paper, we aim at achieving good reconstructed image quality when using low magnetic field and high reduction factor. Under these severe experimental conditions, neither the SENSE method nor the Tikhonov regularization in the image domain give convincing results. To this aim, we present a novel method for SENSE-based reconstruction which proceeds with regularization in the complex wavelet domain. To further enhance the reconstructed image quality, local convex constraints are added in the regularization process. In vivo experiments carried out on Gradient-Echo (GRE) anatomical and Echo-Planar Imaging (EPI) functional MRI data at 1.5 Tesla indicate that the proposed algorithm provides reconstructed images with reduced artifacts for high reduction factor.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。