A difficult task with many particle detectors focusing on interactions below ~100 keV is to perform a calibration in the appropriate energy range that adequately probes all regions of the detector. Because detector response can vary greatly in various locations within the device, a spatially uniform calibration is important. We present a new method for calibration of liquid xenon (LXe) detectors, using the short-lived 83m-Kr. This source has transitions at 9.4 and 32.1 keV, and as a noble gas like Xe, it disperses uniformly in all regions of the detector. Even for low source activities, the existence of the two transitions provides a method of identifying the decays that is free of background. We find that at decreasing energies, the LXe light yield increases, while the amount of electric field quenching is diminished. Additionally, we show that if any long-lived radioactive backgrounds are introduced by this method, they will present less than 67E-6 events/kg/day in the next generation of LXe dark matter direct detection searches
We report the preparation of a Kr-83m source and its subsequent use in calibrating a liquid xenon detector. Kr-83m atoms were produced through the decay of Rb-83 atoms trapped in zeolite molecular sieve and were then introduced into liquid xenon. Decaying Kr-83m nuclei were detected through liquid xenon scintillation. Conversion electrons with energies of 9.4 keV and 32.1 keV from the decay of Kr-83m were both observed. This calibration source will allow the characterization of the scintillation and ionization response of noble liquid detectors at low energies, highly valuable for the search f…
The isomer 83mKr with its half-life of 1.83 h is an ideal calibration source for a liquid noble gas dark matter experiment like the XENON project. However, the risk of contamination of the detector with traces of the much longer lived mother isotop 83Rb (86.2 d half-life) has to be ruled out. In this work the release of 83Rb atoms from a 1.8 MBq 83Rb source embedded in zeolite beads has been investigated. To do so, a cryogenic trap has been connected to the source for about 10 days, after which it was removed and probed for the strongest 83Rb gamma-rays with an ultra-sensitive Germanium detect…
The XENON experimental program aims to detect cold dark matter particles via their elastic collisions with xenon nuclei in two-phase time projection chambers (TPCs). We are currently testing a new TPC at the 100 kg scale, XENON100. This new, ultra-low background detector, has a total of 170 kg of xenon (65 kg in the target region and 105 kg in the active shield). It has been installed at the Gran Sasso Underground Laboratory and is currently in commissioning phase. We review the design and performance of the detector and its associated systems, present status, preliminary calibration results, …
Detectors using liquid xenon as target are widely deployed in rare event searches. Conclusions on the interacting particle rely on a precise reconstruction of the deposited energy which requires calibrations of the energy scale of the detector by means of radioactive sources. However, a microscopic calibration, i.e. the translation from the number of excitation quanta into deposited energy, also necessitates good knowledge of the energy required to produce single scintillation photons or ionisation electrons in liquid xenon. The sum of these excitation quanta is directly proportional to the de…
We characterize two 40 kBq sources of electrodeposited Th-228 for use in low-background experiments. The sources efficiently emanate Rn-220, a noble gas that can diffuse in a detector volume. Rn-220 and its daughter isotopes produce alpha, beta, and gamma-radiation, which may used to calibrate a variety of detector responses and features, before decaying completely in only a few days. We perform various tests to place limits on the release of other long-lived isotopes. In particular, we find an emanation of <0.008 atoms/min/kBq (90% CL) for Th-228 and 1.53 atoms/min/kBq for Ra-224. The sources…
Results of transmission experiments of vacuum ultraviolet light through a 11.6 cm long cell filled with pure and xenon-doped liquid argon are described. Pure liquid argon shows no attenuation down to the experimental short-wavelength cut-off at 118nm. Based on a conservative approach, a lower limit of 1.10 m for the attenuation length of its own scintillation light could be derived. Adding xenon to liquid argon at concentrations on the order of parts per million leads to strong xenon-related absorption features which are used for a tentative assignment of the recently found near-infrared emiss…
Taking liquid xenon detectors as a case study, the importance of a robust recoil energy calibration as a prerequisite to a search for light-mass Weakly Interacting Massive Particles (WIMPs) is emphasized. Important shortfalls in the analysis of existing measurements of the relative scintillation efficiency and ionization yield for nuclear recoils in liquid xenon are described, leading to the conclusion that recent attempts to extract light-WIMP sensitivity limits from the XENON10 and XENON100 detectors are premature and overly optimistic.
We present fore-optics and calibration unit design of Devasthal Optical Telescope Integral Field Spectrograph (DOTIFS). DOTIFS fore-optics is designed to modify the focal ratio of the light and to match its plate scale to the physical size of Integral Field Units (IFUs). The fore-optics also delivers a telecentric beam to the IFUs on the telescope focal plane. There is a calibration unit part of which is combined with the fore-optics to have a light and compact system. We use Xenon-arc lamp as a continuum source and Krypton/Mercury-Neon lamps as wavelength calibration sources. Fore-optics and …
The usual nuclear recoil energy reconstruction employed by liquid xenon dark matter search experiments relies only on the primary scintillation photon signal. Energy reconstruction based on both the photon and electron signals yields a more accurate representation of search results. For a dark matter particle mass m~10 GeV, a nuclear recoil from a scattering event is more likely to be observed in the lower left corner of the typical search box, rather than near the nuclear recoil calibration centroid. In this region of the search box, the actual nuclear recoil energies are smaller than the usu…
The Large Underground Xenon (LUX) dark matter experiment is operating 1.5 km underground at the Sanford Underground Research Facility in Lead, South Dakota, USA. In 2013, the experiment had a WIMP search exposure of 10,091 kg-days over a period of 85.3 live-days. This first dark matter search placed the world's most stringent limits on WIMP-nucleon interaction cross-sections over a wide range of WIMP masses, and is in tension with signal hints of low-mass WIMPs from DAMA, CoGeNT and CDMS-II Si. LUX will commence a 300 day run in 2014 that will improve the sensitivity by a factor of 5. Low-ener…
The radioactive isomer $^{83\mathrm{m}}$Kr has many properties that make it very useful for various applications. Its low energy decay products, like conversion, shake-off and Auger electrons as well as X- and $γ$-rays are used for calibration purposes in neutrino mass experiments and direct dark matter detection experiments. Thanks to the short half-life of 1.83 h and the decay to the ground state $^{83}$Kr, one does not risk contamination of any low-background experiment with long- lived radionuclides. In this paper, we present two new applications of $^{83\mathrm{m}}$Kr. It can be used as a…
The Large Underground Xenon (LUX) experiment is a dual-phase liquid xenon time projection chamber (TPC) operating at the Sanford Underground Research Facility in Lead, South Dakota. A calibration of nuclear recoils in liquid xenon was performed $\textit{in situ}$ in the LUX detector using a collimated beam of mono-energetic 2.45 MeV neutrons produced by a deuterium-deuterium (D-D) fusion source. The nuclear recoil energy from the first neutron scatter in the TPC was reconstructed using the measured scattering angle defined by double-scatter neutron events within the active xenon volume. We mea…
A small-scale, two-phase (liquid/gas) xenon time projection chamber (Xurich II) was designed, constructed and is under operation at the University of Zurich. Its main purpose is to investigate the microphysics of particle interactions in liquid xenon at energies below 50 keV, which are relevant for rare event searches using xenon as target material. Here we describe in detail the detector, its associated infrastructure, and the signal identification algorithm developed for processing and analysing the data. We present the first characterisation of the new instrument with calibration data from …
Direction dependent calibration of widefield radio interferometers estimates the systematic errors along multiple directions in the sky. This is necessary because with most systematic errors that are caused by effects such as the ionosphere or the receiver beam shape, there is significant spatial variation. Fortunately, there is some deterministic behavior of these variations in most situations. We enforce this underlying smooth spatial behavior of systematic errors as an additional constraint onto spectrally constrained direction dependent calibration. Using both analysis and simulations, we …
The XENON100 dark matter experiment uses liquid xenon in a time projection chamber (TPC) to measure xenon nuclear recoils resulting from the scattering of dark matter Weakly Interacting Massive Particles (WIMPs). In this paper, we report the observation of single-electron charge signals which are not related to WIMP interactions. These signals, which show the excellent sensitivity of the detector to small charge signals, are explained as being due to the photoionization of impurities in the liquid xenon and of the metal components inside the TPC. They are used as a unique calibration source to…
We have measured the energy dependence of the liquid xenon (LXe) scintillation yield of electrons with energy between 2.1 and 120.2keV, using the Compton coincidence technique. A LXe scintillation detector with a very high light detection efficiency was irradiated with 137Cs γ rays and the energy of the Compton-scattered γ rays was measured with a high-purity germanium (HPGe) detector placed at different scattering angles. The excellent energy resolution of the HPGe detector allows the selection of events with Compton electrons of known energy in the LXe detector. We find that the scintillatio…
Results from the nuclear recoil calibration of the XENON100 dark matter detector installed underground at the Laboratori Nazionali del Gran Sasso (LNGS), Italy are presented. Data from measurements with an external 241AmBe neutron source are compared with a detailed Monte Carlo simulation which is used to extract the energy dependent charge-yield Qy and relative scintillation efficiency Leff. A very good level of absolute spectral matching is achieved in both observable signal channels - scintillation S1 and ionization S2 - along with agreement in the 2-dimensional particle discrimination spac…
We studied the application of statistical reconstruction algorithms, namely maximum likelihood and least squares methods, to the problem of event reconstruction in a dual phase liquid xenon detector. An iterative method was developed for in-situ reconstruction of the PMT light response functions from calibration data taken with an uncollimated gamma-ray source. Using the techniques described, the performance of the ZEPLIN-III dark matter detector was studied for 122 keV gamma-rays. For the inner part of the detector (R<100 mm), spatial resolutions of 13 mm and 1.6 mm FWHM were measured in the …
We report an absolute calibration of the ionization yields($\textit{Q$_y$})$ and fluctuations for electronic recoil events in liquid xenon at discrete energies between 186 eV and 33.2 keV. The average electric field applied across the liquid xenon target is 180 V/cm. The data are obtained using low energy $^{127}$Xe electron capture decay events from the 95.0-day first run from LUX (WS2013) in search of Weakly Interacting Massive Particles (WIMPs). The sequence of gamma-ray and X-ray cascades associated with $^{127}$I de-excitations produces clearly identified 2-vertex events in the LUX detect…
Bubble chambers using liquid xenon (and liquid argon) have been operated (resp. planned) by the Scintillating Bubble Chamber (SBC) collaboration for GeV-scale dark matter searches and CE$ν$NS from reactors. This will require a robust calibration program of the nucleation efficiency of low-energy nuclear recoils in these target media. Such a program has been carried out by the PICO collaboration, which aims to directly detect dark matter using $\mathrm{C_3 F_8}$ bubble chambers. Neutron calibration data from mono-energetic neutron beam and Am-Be source has been collected and analyzed, leading t…
We report results from an extensive set of measurements of the \b{eta}-decay response in liquid xenon.These measurements are derived from high-statistics calibration data from injected sources of both $^{3}$H and $^{14}$C in the LUX detector. The mean light-to-charge ratio is reported for 13 electric field values ranging from 43 to 491 V/cm, and for energies ranging from 1.5 to 145 keV.
Calibration of radio interferometric observations becomes increasingly difficult towards lower frequencies. Below ~300 MHz, spatially variant refractions and propagation delays of radio waves traveling through the ionosphere cause phase rotations that can vary significantly with time, viewing direction and antenna location. In this article we present a description and first results of SPAM (Source Peeling and Atmospheric Modeling), a new calibration method that attempts to iteratively solve and correct for ionospheric phase errors. To model the ionosphere, we construct a time-variant, 2-dimens…
With its array of 3840 Transition Edge Sensors (TESs) operated at 90 mK, the X-Ray Integral Field Unit (X-IFU) on board the ESA L2 mission Athena will provide spatially resolved high-resolution spectroscopy (2.5 eV FWHM up to 7 keV) over the 0.2 to 12 keV bandpass. The in-flight performance of the X-IFU will be strongly affected by the calibration of the instrument. Uncertainties in the knowledge of the overall system, from the filter transmission to the energy scale, may introduce systematic errors in the data, which could potentially compromise science objectives - notably those involving li…
The Herschel SPIRE instrument consists of an imaging photometric camera and an imaging Fourier Transform Spectrometer (FTS), both operating over a frequency range of 450-1550 GHz. In this paper, we briefly review the FTS design, operation, and data reduction, and describe in detail the approach taken to relative calibration (removal of instrument signatures) and absolute calibration against standard astronomical sources. The calibration scheme assumes a spatially extended source and uses the Herschel telescope as primary calibrator. Conversion from extended to point-source calibration is carri…
LUX, the world's largest dual-phase xenon time-projection chamber, with a fiducial target mass of 118 kg and 10,091 kg-days of exposure thus far, is currently the most sensitive direct dark matter search experiment. The initial null-result limit on the spin-independent WIMP-nucleon scattering cross-section was released in October 2013, with a primary scintillation threshold of 2 phe, roughly 3 keVnr for LUX. The detector has been deployed at the Sanford Underground Research Facility (SURF) in Lead, South Dakota, and is the first experiment to achieve a limit on the WIMP cross-section lower tha…
We present a high-precision measurement of the parallax for the 12-day Cepheid SS Canis Majoris, obtained via spatial scanning with the Wide Field Camera 3 (WFC3) on the Hubble Space Telescope (HST). Spatial scanning enables astrometric measurements with a precision of 20-40 muas, an order of magnitude better than pointed observations. SS CMa is the second Cepheid targeted for parallax measurement with HST, and is the first of a sample of eighteen long-period >~ 10 days) Cepheids selected in order to improve the calibration of their period-luminosity relation and eventually permit a determinat…
We present measurements of the electron-recoil (ER) response of the LUX dark matter detector based upon 170,000 highly pure and spatially-uniform tritium decays. We reconstruct the tritium energy spectrum using the combined energy model and find good agreement with expectations. We report the average charge and light yields of ER events in liquid xenon at 180 V/cm and 105 V/cm and compare the results to the NEST model. We also measure the mean charge recombination fraction and its fluctuations, and we investigate the location and width of the LUX ER band. These results provide input to a re-an…