arXivDaily arXiv每日学术速递 周一至周五更新
arXiv 0903.3398quant-ph
收藏

用于灵敏脉冲检测应用的超低噪声差分交流耦合光电探测器

Ultra--low noise differential AC-coupled photodetector for sensitive pulse detection applications

P. J. Windpassinger, M. Kubasik, M. Koschorreck, A. Boisen, N. Kjaergaard, E. S. Polzik, J. H. Mueller

摘要:

我们报告的性能超低噪声差分光电探测器,特别是设计用于探测原子系综与弱光脉冲。介绍了探测器的工作原理,以及用于提取具有$\sim1 μ$持续时间的光脉冲的光子散粒噪声的分析方法。与频率响应峰值检测器相反,我们的方法允许宽带量子噪声测量。两种不同硬件方法的等效噪声电荷(ENC)分别被评估为每脉冲280和340个电子,其对应于两种方法中的$n_\mathrm{3dB}=0.8\cdot 10^5$和$n_\mathrm{3dB}=1.2\cdot 10^5$的暗噪声等效光子数。最后,我们讨论了用双相关采样方法消除由探测器缺陷引起的输出信号中的经典相关的可能性。

英文摘要:

We report on the performance of ultra low noise differential photodetectors especially designed for probing of atomic ensembles with weak light pulses. The working principle of the detectors is described together with the analysis procedures employed to extract the photon shot noise of light pulses with $\sim1 μ$s duration. As opposed to frequency response peaked detectors, our approach allows for broadband quantum noise measurements. The equivalent noise charge (ENC) for two different hardware approaches is evaluated to 280 and 340 electrons per pulse, respectively which corresponds to a dark noise equivalent photon number of $n_\mathrm{3dB}=0.8\cdot 10^5$ and $n_\mathrm{3dB}=1.2\cdot 10^5$ in the two approaches. Finally, we discuss the possibility of removing classical correlations in the output signal caused by detector imperfection by using double--correlated sampling methods.