摘要:
一个p型HPGe探测器已被分割切割,用金刚石锯,并蚀刻六个圆周槽通过锂扩散死层。分割的程度进行了测试与88千电子伏的伽马射线从一个良好的准直源的109镉。被识别为发生在多于一个段中并且从能量间隔2038 +/-5 keV被拒绝的事件的分数被测量为0.59。
英文摘要:
A p-type HPGe detector has been segmented by cutting, with a diamond saw, and etching six circumferential grooves through the Li-diffused dead layer. The degree of segmentation was tested with the 88 keV gamma rays from a well-collimated source of 109Cd. The fraction of events, recognized as occurring in more than one segment, and rejected from the energy interval 2038 +/- 5 keV, was measured as 0.59.