双端口网络分析仪测量中的总微分误差

Total Differential Errors in Two-Port Network Analyser Measurements

N. I. Yannopoulou, P. E. Zimourtopoulos

摘要:

由于没有不确定性的S参数测量不能声称任何可信度,在全双端口矢量网络分析仪(VNA)测量的不确定性估计使用总复微分(总微分误差)。为了精确地表达复微分误差之间的比较关系,使用了它们的微分误差区域(DER)。为了证明该方法在最准确的情况下,直接零长度通过,实际结果是常用的Z参数的一个简单的,两个端口,直流电阻T-网络,这是建立和测试对频率与VNA测量系统扩展了两个漫长的传输线。

英文摘要:

Since S-parameter measurements without uncertainty cannot claim any credibility, the uncertainties in full two-port Vector Network Analyser (VNA) measurements were estimated using total complex differentials (Total Differential Errors). To express precisely a comparison relation between complex differential errors, their differential error regions (DERs) were used. To demonstrate the method in the most accurate case of a direct zero-length thru, practical results are presented for commonly used Z-parameters of a simple, two-port, DC resistive T-network, which was built and tested against frequency with a VNA measurement system extended by two lengthy transmission lines.