手动、操纵杆还是触觉控制?机器人介入神经放射学程序导航策略的体外比较

Manual, Joystick, or Haptic Control? An In Vitro Comparison of Navigation Strategies for Robotic Interventional Neuroradiology Procedures

School of Biomedical Engineering & Imaging Sciences, King’s College London, UK(伦敦国王学院生物医学工程与成像科学学院);School of Engineering and Materials Science, Queen Mary University of London, UK(伦敦女王学院工程与材料科学学院);Department of Neuroradiology, King’s College Hospital NHS Foundation Trust, London, UK(伦敦国王学院医院国家健康服务信托基金会神经放射科);The Walton Centre NHS Foundation Trust, Liverpool, UK(利物浦沃尔顿中心国家健康服务信托基金会);University College London Hospitals NHS Foundation Trust, London, UK(伦敦大学学院医院国家健康服务信托基金会);East Kent Hospitals University NHS Foundation Trust, Kent, UK(肯特郡东肯特医院大学国家健康服务信托基金会)
Benjamin Jackson, Nikola Fischer, Harry Robershaw, Xingyu Chen, S. H. Hadi Sadati, Yang Li, Jeremy Lynch, Nasr Abdelsalam, Jonathon Buwanabala, Matthew Benger, … 展开作者

Benjamin Jackson, Nikola Fischer, Harry Robershaw, Xingyu Chen, S. H. Hadi Sadati, Yang Li, Jeremy Lynch, Nasr Abdelsalam, Jonathon Buwanabala, Matthew Benger, Sara Sciacca, Naga Kandasamy, Marco Mancuso-Marcello, Parthiban Balasundaram, Sahan Guruge, Neelan Das, Alejandro Granados, Kawal Rhode, Thomas C Booth

AI总结:

该研究比较了机器人介入神经放射学程序中手动、操纵杆和触觉控制三种导航策略,开发相关设备和模型,让10名医生用四种方式模拟导航,评估多项指标。结果显示手动导航最快,各方式均安全,操纵杆脱垂事件多,专家表现更好,触觉反馈有潜力。

摘要:

目的:评估用于介入神经放射学程序的机器人控制器接口,该接口结合力传感平台以评估安全性。方法:开发了定制的血管内机器人、设备模拟控制器和传感神经血管模型。10名介入神经放射科医生(4名新手,6名专家)使用四种控制方式进行模拟导航:有无触觉反馈的设备模拟控制器、基于操纵杆的输入和手动导航。评估了导航时间、血管壁峰值力、错误插管和脱垂事件以及用户分析。结果:与触觉开启(248.7秒)、触觉关闭(314.7秒)和操纵杆(392.6秒)方式相比,手动导航最快(平均47.7秒)(p<0.001)。无论控制器类型如何,血管壁力均低于0.70N穿刺阈值;因此所有方式都被认为是安全的。操纵杆产生的脱垂事件明显多于手动控制(1.56对0.13;p=0.018)。操作者经验与性能相关:专家的错误插管比新手少(0.25对0.62;p=0.035),施加的血管壁力也较小(p<0.0005);这些影响在所有控制器中都存在,但在触觉开启时更为明显。用户认为触觉开启和触觉关闭同样直观,比操纵杆更直观(p=0.033)。结论:在大多数指标上,设备模拟机器人控制器优于操纵杆接口;触觉反馈显示出有希望但不显著的性能优势。

英文摘要:

Objective: To evaluate robotic controller interfaces for interventional neuroradiology procedures in-vitro incorporating a force-sensing platform to assess safety. Methods: A custom endovascular robot, device-mimicking controller, and sensorized neurovascular phantom were developed. Ten interventional neuroradiologists (4 novices, 6 experts) performed simulated navigations using four control modalities: device-mimicking controllers with and without haptic feedback, joystick-based input, and manual navigation. Navigation time, peak vessel-wall forces, incorrect catheterisations, and prolapse events were assessed, alongside user analyses. Results: Manual navigation was fastest (mean 47.7 s) compared to haptic-on (248.7 s), haptic-off (314.7 s), and joystick (392.6 s) modalities (p<0.001). Regardless of controller type, vessel-wall forces were below the 0.70 N puncture threshold; therefore all modalities were considered safe. Joystick produced significantly more prolapse events than manual control (1.56 vs 0.13; p=0.018). Operator experience was relevant to performance: experts made fewer incorrect catheterisations than novices (0.25 vs 0.62; p=0.035) and applied less vessel-wall force (p<0.0005); these effects were sustained across controllers but accentuated when haptics were on. Users perceived haptic on and haptic off as similarly intuitive, and more intuitive than joystick (p=0.033). Conclusion: Device-mimicking robotic controllers outperform joystick interfaces on most metrics; haptic feedback shows promising but non-significant performance benefits.