基于梯度计的电磁定位系统实现腔内医疗器械的闭式位姿估计

Closed-Form 6-DoF Pose Estimation of Medical Tools via Gradiometer-Based Electromagnetic Localization System

Zhiwei Wu, Jiahao Luo, Yubo Pu, Siyi Wei, Yuankai Chen, Jinhui Zhang

AI总结:

提出一种基于梯度计的电磁定位系统(GELS),利用紧凑型磁力计阵列作为准梯度计估计局部磁场和梯度张量,通过欧拉齐次关系映射为位移,再经多源Procrustes配准实现闭式位姿估计,无需预校准场图或迭代优化。

摘要:

嵌入式磁跟踪对于腔内医疗器械的远程导航具有极具吸引力的前景。然而,现有的六自由度位姿恢复方法通常需要预校准的工作空间场图或迭代非线性优化。本文提出了一种基于梯度计的电磁定位系统(GELS),这是一种闭式跟踪框架,使用紧凑型磁力计阵列作为嵌入式准梯度计来估计局部磁场和梯度张量。这些量通过欧拉齐次关系映射为源与阵列之间的位移,随后利用至少三个非共线源的多源Procrustes配准恢复阵列的方向和位置。该算法需要已知的源位置和阵列几何结构,但无需预校准的工作空间场图、初始位姿猜测或校准的激励源矩。恢复的位姿还可作为移动磁参考框架,实现概念验证的子级偶极子定位任务。跨传感器阵列配置和激励模式的台架实验显示,序列平均位置误差为\SI{10.80}{\milli\meter}--\SI{15.57}{\milli\meter},最快更新率为\SI{14.49}{\hertz},中位求解器运行时间为\SI{172.00}{\micro\second}。基于扰动的误差传播分析进一步确定了传感器间不一致性和偶极子模型失配是主要的精度限制因素,从而为未来进一步减少位姿估计误差的传感器阵列和磁源设计提供指导。

英文摘要:

Electromagnetic tracking \added[]{is promising for} remote navigation of endoluminal \added[]{medical tools}. However, existing six-degree-of-freedom (6-DoF) pose recovery approaches often require pre-calibrated workspace field maps or iterative nonlinear optimization. This letter presents a Gradiometer-Based Electromagnetic Localization System (GELS), a closed-form tracking framework that uses a compact magnetometer array as an embedded quasi-gradiometer to estimate local magnetic fields and gradient tensors. These quantities are mapped by the Euler homogeneous relation to displacements between source and array, from which Procrustes registration over multiple sources recovers the array orientation and position using at least three noncollinear sources. The algorithm requires known source positions and array geometry, but no pre-calibrated workspace field maps, initial pose guesses, or calibrated source magnetic moments. The recovered \added[]{array} pose also enables a demonstration of sublevel dipole localization by serving as a mobile magnetic reference frame. Benchtop experiments across sensor array configurations and excitation modes validate the proposed closed-form 6-DoF localization pipeline, with a median solver runtime of \SI{172.00}{\micro\second}. A perturbation analysis identifies inter-sensor inconsistency and dipole model mismatch as dominant accuracy limits, informing future sensor array and source design to improve pose accuracy.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑