基于法布里-珀罗的PARAS-2光谱仪波长校准系统的设计与开发
Design and development of Fabry-Perot based wavelength calibration system for PARAS-2 spectrograph
AI总结:
针对高精度径向速度光谱仪的波长校准需求,开发了一种基于法布里-珀罗标准具的低成本波长校准器,与氙弧灯结合集成到PARAS-2光谱仪中,通过温压控制产生密集梳状光谱,实现10 cm/s的理论稳定性和40-70 cm/s的实测相对漂移。
摘要:
精确的波长校准对于高精度径向速度(RV)光谱仪至关重要,需要一种稳定的校准器,提供密集且均匀间隔的谱线网格,以准确确定恒星谱线位置并监测仪器漂移。在这项工作中,我们介绍了一种基于法布里-珀罗(FP)标准具的低成本波长校准器的开发,旨在克服空心阴极灯(HCL)和碘池等传统光源的局限性。该FP校准器与氙(Xe)弧灯组件结合,已集成到位于阿布山天文台PRL 2.5米望远镜上的PARAS-2光谱仪中。在受控温度和压力条件下运行,该系统生成覆盖62个阶梯光栅级次、包含超过10,000条清晰稳定谱线的密集梳状光谱,从而实现仪器漂移的精确测量。初步结果显示,自由光谱范围(FSR)从4000 Å附近的0.16 Å变化到7000 Å附近的0.49 Å,在中心波长5500 Å处约为0.3 Å。估计的精细度从4000 Å附近的9变化到6900 Å附近的19,在5500 Å处约为17。温度和压力稳定性测试显示RMS变化分别为$0.002 ^\circ\mathrm{C}$和$5 imes10^{-4}$ mbar。基于这些值,FP波长校准器的理论稳定性估计在10 cm/s以内,使其成为高分辨率光谱校准中激光频率梳(LFC)的可靠替代方案。我们提供了FP校准器RV稳定性的初步评估,得到40-70 cm/s的相对漂移,这有待进一步研究。观察到的超出理论估计限值的过量可能归因于氙弧灯中电弧漂移引起的不稳定性。
英文摘要:
Precise wavelength calibration is essential for high-precision radial velocity (RV) spectrographs, necessitating a stable calibrator that provides a dense grid of uniformly spaced lines to accurately determine stellar line positions and monitor instrumental drifts. In this work, we present the development of a cost-effective Fabry-Perot (FP) etalon-based wavelength calibrator designed to overcome the limitations of conventional sources such as hollow cathode lamps (HCLs) and iodine cells. This FP calibrator, combined with a Xenon (Xe) arc lamp assembly, has been integrated with the PARAS-2 spectrograph on the PRL 2.5m telescope at Mount Abu Observatory. Operated under controlled temperature and pressure conditions, the system generates a dense, comb-like spectrum covering 62 echelle orders with more than 10,000 well-defined and stable spectral lines, enabling precise measurement of instrumental drift. Initial results show that the free spectral range (FSR) varies from 0.16 Å~near 4000 Å~to 0.49 Å~ near 7000 Å, with a value of 0.3 Å~around the central wavelength of 5500 Å~. The estimated finesse ranges from 9 near 4000 Å~to 19 near 6900 Å, with an approximate value of 17 at 5500 Å. The temperature and pressure stability tests demonstrate RMS variations of $0.002 ^\circ\mathrm{C}$ and $5\times10^{-4}$ mbar, respectively. Based on these values, the theoretical stability of the FP wavelength calibrator is estimated to be within 10 cm/s, establishing it as a reliable alternative to Laser Frequency Combs (LFCs) for high-resolution spectroscopic calibration. We present an initial assessment of the RV stability of the FP calibrator, yielding 40-70 cm/s of relative drifts, which are up for further investigations. The observed excess over the theoretically estimated limit is likely attributable to instabilities arising from arc wandering in the xenon arc lamp.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。