研究用于双向仿生接口的无线电力传输问题,设计了低频(127kHz)感应WPT链路,含自谐振驱动控制。通过选择频率等设计无线体域传感系统,原型能提供约140mA和约20V,效率超40%,满足2cm植入深度要求。

摘要:

为实现多模式感官反馈和运动控制,先进假肢系统的双向仿生接口需要向植入式电子设备持续且安全地供电,并集成到患者的传感无线体域网中。然而,为这类接口供电仍是一个未解决的问题。无线电力传输(WPT)避免了植入电池和经皮连接,但其设计受到电磁安全、植入尺寸、电压兼容性以及与敏感生物信号采集和刺激电路共存等严格要求的限制。本文介绍了一种用于植入式双向仿生接口的低频(127kHz)感应WPT链路的设计与测试。该系统包括自谐振驱动控制,以在网络物理假体的耦合和负载条件变化时保持共振运行。从仿生接口的要求出发,通过选择工作频率、绘制电气方案并检查其安全性和法规合规性,设计了无线体域传感系统。初步的WPT原型可提供高达约140mA和约20V的电压,实现了高于40%的最大功率传输效率,并满足了高达2cm植入深度的项目要求。

英文摘要:

To provide multimode sensory feedback and motion control, bidirectional bionic interfaces for advanced prosthetic systems require continuous and secure energy delivery to implantable electronics and integration in the sensing WBAN (Wireless Body Area Network) of the patient. However, powering such interfaces is still an open issue. Wireless Power Transfer (WPT) avoids implanted batteries and transcutaneous connections, but its design is constrained by stringent requirements on electromagnetic safety, implant size, voltage compliance, and coexistence with sensitive bio-signal acquisition and stimulation circuitry. This paper presents the design and testing of a low-frequency (127 kHz) inductive WPT link for an implantable bidirectional bionic interface. The system includes an autoresonant driving control to maintain operation at resonance under varying coupling and load conditions of the cyber-physical prosthesis. Starting from the requirements of the bionic interface, the wireless body-area sensing system is designed by selecting the working frequency, drawing the electrical schemes, and checking its safety and regulatory compliance. Preliminary WPT prototypes can provide up to ~140 mA and ~20 V, achieving a maximum power transfer efficiency higher than 40% and satisfying the project requirements up to a 2 cm implantation depth.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑