罗伯特·斯托比光谱仪上的狭缝掩模积分场单元:I. 仪器开发

A Slit Mask Integral Field Unit for the Robert Stobie Spectrograph on the Southern African Large Telescope: I. Instrument Development

Space Science and Engineering Initiative, College of Engineering, University of Hawai’i at Manoa(夏威夷大学马诺阿分校工程学院空间科学与工程倡议);Centre for Space Research, North-West University(西北大学太空研究中心);South African Astronomical Observatory(南非天文台);University of Wisconsin(威斯康星大学)

Sabyasachi Chattopadhyay, Matthew A. Bershady, Michael P. Smith, Antoine Mahoro, Marsha J. Wolf

AI总结:

本文介绍了为南大西洋大望远镜罗伯特·斯托比光谱仪开发的积分场单元,采用光纤和棱镜折叠镜构建紧凑型设备,实现高通量和高分辨率的光谱观测。

摘要:

积分场光谱学(IFS)作为新的观测模式被添加到罗伯特·斯托比光谱仪(RSS)中,这是南大西洋大望远镜上的多模式仪器。RSS作为成像光谱仪覆盖320-900 nm,具有6600 arcsec的光谱分辨率-狭缝宽度乘积。利用光纤和棱镜折叠镜,我们能够构建紧凑的积分场单元(IFUs),其体积与长狭缝机匣(134 mm x 130 mm x 8 mm)相同。这些'狭缝掩模'IFUs(SMIs)将望远镜光束导向面向天空的2D光纤阵列,该阵列在焦点平面维度上路由成一个8′1D伪狭缝,光纤输出被重新导向光谱仪聚光器。第一个完成的单元SMI-200具有303个目标光纤和24个天空光纤,提供0.8′(200 μm核心直径)的空间分辨率,覆盖22.5′x 17.6′的视场。本文描述了特定的设计考虑和设计与制造策略,以在严苛且高度受限的机匣几何条件下最大化性能并最小化风险。我们还详细说明了IFU的映射和实验室表征。实验室测量显示总通量为77%,但在RSS聚光器接受光束(f/4.2)内由于由紧密机匣体积引起的锐弯半径引起的损耗,有效通量仅为55-60%。

英文摘要:

Integral field spectroscopy (IFS) has been added as a new observation mode to the Robert Stobie Spectrograph (RSS), the workhorse multi-mode instrument on the Southern African Large Telescope. RSS operates as an imaging spectrograph covering 320-900 nm with a spectral resolution--slit-width product of 6600 arcsec. Using fiber optics and prismatic fold mirrors, we have been able to construct compact integral field units (IFUs) that fit within the same volume as the long-slit cassettes (134 mm x 130 mm x 8 mm). These `slit mask' IFUs (SMIs) direct the telescope beam into a 2D sky-facing fiber array routed in the focal plane dimension into an 8$\arcmin$ 1D pseudo-slit, with fiber output redirected back into the spectrograph collimator. The first completed unit, SMI-200, features 303 object fibers and 24 sky fibers, providing a spatial resolution of 0.8$\arcsec$ (200 $\upmu$m core diameter) over a field of view (FOV) of 22.5$\arcsec$ x 17.6$\arcsec$. This paper describes the specific design considerations and design and fabrication strategies to maximize performance and minimize risk during construction, given the demanding and highly constrained cassette geometry. We also detail mapping and laboratory characterization of the IFU. Laboratory measurements demonstrate a total throughput of 77\%, but an effective throughput of only 55-60\% within the RSS collimator acceptance beam of f/4.2 due to losses dominated by focal ratio degradation (FRD) induced by sharp bend radii imposed by the tight cassette volume.