太平洋中微子实验的光学校准系统
Optical calibration systems of the Pacific Ocean Neutrino Experiment
AI总结:
本文介绍P-ONE实验的光学校准系统,包括基于GaN FET的脉冲驱动电路和定向/各向同性校准模块,实现增益、能量和时间校准,并展示性能表征结果。
摘要:
本文介绍了为太平洋中微子实验(P-ONE)生产的光学校准系统的设计和性能表征,这些系统针对探测器中的增益、能量和时间校准。这些系统包括基于氮化镓场效应晶体管技术的新型光脉冲驱动电路及其在定向和各向同性自监测光学校准仪器中的应用。为第一条P-ONE线生产了总共330个定向光脉冲器和两个各向同性的17英寸校准模块(P-CAL)。我们展示了定向和各向同性校准装置的设计和性能,并对两个完整生产批次进行了详细的光学表征。在$365 - 520\,$nm波长范围内,我们开发的驱动电路分别实现了高达$10^{11}\,$光子的发射强度和低至$1.4\,$ns的脉冲宽度。P-CAL中的光脉冲驱动器和自监测电子器件使用相同的实验装置进行了表征,并结合专用的基于GEANT4的模拟框架优化了仪器的光学各向同性设计。优化后的P-CAL在整个$4\pi\,$立体角范围内实现了$1.00 \pm 0.01$的模拟各向同性等级。这些模拟研究通过使用两个独立实验装置在空气和水中进行的专门测量得到了明确确认,我们报告了结果。由此,可以对P-ONE中部署的P-CAL模块进行详细的性能估计。
英文摘要:
This work presents the design and performance characterization of the optical calibration systems produced for the Pacific Ocean Neutrino Experiment (P-ONE), which target gain, energy and time calibration in the detector. These systems include novel light-pulse driver circuitry based on gallium nitride field-effect transistor technology and its application to directional and isotropic, self-monitoring optical calibration instruments. A total of 330 directional light pulsers and two isotropic, 17-inch calibration modules (P-CALs) were produced for the first P-ONE line. We present the designs and performance of both the directional and isotropic calibration devices and perform detailed optical characterizations of both full-production batches. In a wavelength range of $365 - 520\,$nm, our developed driver circuits achieve emission intensities up to $10^{11}\,$photons and pulse widths as small as $1.4\,$ns, respectively. Light-pulse drivers and self-monitoring electronics in the P-CAL were characterized using the same experimental setup, and the instrument's optical-isotropy design was optimized in combination with a dedicated GEANT4-based simulation framework. The optimized P-CAL achieves a simulated isotropy grade of $1.00 \pm 0.01$ across the entire $4π\,$solid angle range. These simulation investigations were explicitly confirmed by dedicated measurements in both air and water using two independent experimental setups, and we report the results. With this, a detailed performance estimate for deployed P-CAL modules in P-ONE was possible.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。