使用超高能量分辨率磁微 calorimeter 测量绝对伽马射线能量

Sub-eV uncertainties on gamma-ray energies using a magnetic microcalorimeter - Potential new reference energies

Université Paris-Saclay(巴黎萨克雷大学);CEA, List, Laboratoire National Henri Becquerel (LNE-LNHB)(法国原子能和替代能源委员会,列表研究所,亨利·贝克勒尔国家实验室);Institute of Micro- and Nanoelectronic Systems (IMS), Karlsruhe Institute of Technology (KIT)(卡尔斯鲁厄理工学院微纳电子系统研究所);Institute for Data Processing and Electronics (IPE), Karlsruhe Institute of Technology (KIT)(卡尔斯鲁厄理工学院数据处理与电子学研究所);GIP ARRONAX(阿隆纳克斯政府利益集团);Laboratoire Subatech, UMR 6457, IMT Nantes Atlantique/CNRS-IN2P3/Nantes(Subatech实验室,UMR 6457,南特大西洋高等矿业学院/法国国家科学研究中心-法国国家核物理与粒子物理研究所/南特大学)

Matias Rodrigues, Mostafa L. Zahir, Martin Loidl, Lucille Chambon, Quentin Drenne, Michael Müller, Sebastian Kempf, Etienne Nigron, Ferid Haddad

AI总结:

利用超高能量分辨率磁微 calorimeter 提高伽马射线能量测量精度,实现部分能量的绝对不确定度降低至0.13 eV。

摘要:

我们展示了使用高能量分辨率磁微 calorimeter 进行能谱学测量得到的27个伽马射线能量,范围从14 keV到136 keV。该光谱仪有八个像素,其在整个能量范围内实现了15至30 eV的能量分辨率。它面对一个低温源采样器,配有四个可移动的源。每个源包含多种放射性核素,包括用于校准光谱仪和校正其非线性的Yb-169和Co-57。通过Co-57、Cd-109、Ba-133、Gd-153、Eu-154、Eu-155、Yb-169、Tm-170、Pb-210、Np-239、Am-241和Am-243的衰变发射的伽马射线能量已重新评估。所获得的最低绝对不确定度为0.13 eV,在105.3 keV处,对应相对不确定度为1.3 ppm。在27个测量的伽马射线中,19个能量的不确定度比现有文献中的有所改进:10个能量的不确定度降低了5倍,4个能量的不确定度降低了一个数量级以上。与通过波长分散光谱法获得的能量有良好的一致性。这项工作显著优于之前通过半导体探测器进行能散光谱法获得的伽马能量,得益于磁微 calorimeter 的超高能量分辨率。

英文摘要:

We present new measurements of 27 gamma-ray energies ranging from 14 keV to 136 keV, obtained using magnetic microcalorimeters for energy-dispersive spectrometry. The spectrometer has eight pixels and achieves a high energy resolution between 15 eV and 30 eV across the entire energy range. It faces a cryogenic source sampler with four movable sources, each containing a mixture of radionuclides, including $^{169}\mathrm{Yb}$ and $^{57}\mathrm{Co}$. Four gamma-ray lines from $^{169}\mathrm{Yb}$and one from 57Co were used to calibrate the spectrometer and correct for its non-linearity in the energy range 14 keV - 130 keV. The gamma-ray energies, emitted through the decay of $^{57}\mathrm{Co}$, $^{109}\mathrm{Cd}$, $^{133}\mathrm{Ba}$, $^{153}\mathrm{Gd}$, $^{154}\mathrm{Eu}$, $^{155}\mathrm{Eu}$, $^{169}\mathrm{Yb}$, $^{170}\mathrm{Tm}$, $^{210}\mathrm{Pb}$, $^{239}\mathrm{Np}$, $^{241}\mathrm{Am}$ and $^{243}\mathrm{Am}$, have been reassessed. This work significantly reduces the uncertainties upon previously reported gamma energies obtained by energy-dispersive spectrometry using semiconductor detectors: the lowest absolute uncertainty achieved is 0.13 eV at 105.3 keV, which corresponds to a relative uncertainty of 1.3 ppm. With respect to the available literature, among the 27 measured gamma-ray lines, the uncertainties of 10 energies were reduced by a factor of 5, while those of 4 additional energies were reduced by more than one order of magnitude.