超越千分之一能量分辨率的AlMn TES探测器
Beyond One-Thousandth Energy Resolution with an AlMn TES Detector
AI总结:
本研究开发了环形AlMn TES用于X射线检测,实现了12.1 eV的FWHM能量分辨率,首次达到0.1%以下,展示了其在高分辨率X射线检测中的潜力。
摘要:
超导转变边传感器(TES)是下一代X射线光谱仪的关键技术,以其卓越的能量分辨率而闻名。在过去十年中,基于AlMn合金薄膜的TES已被广泛应用于几个宇宙微波背景(CMB)实验中。简单的制造工艺和易于调节的临界温度使其成为双层TES的替代品。然而,直到现在,它们很少被应用于X射线检测。我们开发了一个环形AlMn TES用于X射线检测,并在稀释制冷机中用超导量子干涉仪(SQUID)放大器进行了测试,实现了在17.48 keV处全宽半高(FWHM)为12.1 ± 0.3 eV。据我们所知,这是首次展示AlMn TES实现能量分辨率低于0.1%的演示,突显了其在高分辨率X射线检测中的潜力。
英文摘要:
The superconducting Transition-Edge Sensor (TES) is a critical technology for next-generation X-ray spectrometers, known for its exceptional energy resolution. In the last decade, TESs based on AlMn alloy films have been extensively used in several cosmic microwave background (CMB) experiments. The advantages of simple fabrication process and easily tunable critical temperature make them an alternative to bilayer TESs. However, they have rarely been applied to X-ray detection until now. We developed an annular AlMn TES for X-ray detection and tested it in a dilution refrigerator with a Superconducting Quantum Interference Device (SQUID) amplifier, achieving an Full Width at Half Maximum (FWHM) of 12.1 +- 0.3 eV at 17.48 keV. To the best of our knowledge, this is the first demonstration of an AlMn TES achieving an energy resolution below 0.1%, highlighting its potential for high-resolution X-ray detection.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。