利用差分CR-39核 track探测器确定慢中子探测效率和背景甄别在混合辐射场中

Experimental Determination of Slow-Neutron Detection Efficiency and Background Discrimination in Mixed Radiation Fields Using Differential CR-39 Track Detectors

Indian Institute of Technology Kanpur(坎普尔印度理工学院)

Ankit Kumar, Tushar Verma, Pankaj Jain, Raj Ganesh Pala, K. P. Rajeev

AI总结:

本文提出利用差分CR-39探测器测量慢中子通量的方法,通过对比涂硼与未涂探测器的轨迹差异,确定慢中子探测效率及背景甄别。

摘要:

在包含带电粒子和快中子的辐射场中测量慢中子很困难,尤其是当辐射组成在事先未知时。本文提出了一种经过测试的方法,利用CR-39固态核 track探测器来测量慢中子通量。两个探测器被一起使用:一个涂有硼的CR-39探测器和一个未涂的CR-39探测器。未涂的探测器记录带电粒子和快中子的轨迹,但不响应慢中子。涂有硼的探测器还检测当慢中子与硼反应产生锂和α粒子时产生的带电粒子。通过从未涂探测器的轨迹密度中减去涂硼探测器的轨迹密度,可以得到可靠且保守的慢中子测量。使用参考热中子源的实验表明,两个探测器之间的差异随暴露时间线性增加。统计分析给出慢中子等效轨迹率约为5.84±0.18个每分钟,明显不同于零。该响应的斜率用于确定涂硼探测器的探测效率。未涂探测器测量由源中快中子泄漏引起的背景。这些结果表明,涂硼的CR-39探测器不能单独用于准确的慢中子测量。可靠的中子通量确定需要同时使用未涂探测器。两个探测器之间的差异提供了一个正确的热中子通量估计,在混合辐射场中以及传统中子探测器无法使用的情况下。

英文摘要:

Measuring slow neutrons is difficult when the radiation field also contains charged particles and fast neutrons, especially when the radiation composition is not known in advance. In this work, we present a tested method to measure slow neutron fluence using CR-39 solid state nuclear track detectors. Two detectors are used together: a boron coated CR-39 detector and an uncoated CR-39 detector.The uncoated detector records tracks from charged particles and fast neutrons but does not respond to slow neutrons. The boron coated detector additionally detects charged particles produced when slow neutrons react with boron and generate lithium and alpha particles. Subtracting the track density of the uncoated detector from that of the boron coated detector provides a reliable and conservative measure of slow neutrons.Experiments using a reference thermal neutron source show that the difference between the two detectors increases linearly with exposure time. Statistical analysis gives a slow neutron equivalent track rate of 5.84 plus or minus 0.18 tracks per minute, clearly different from zero. The slope of this response is used to determine the detection efficiency of the boron coated detector. The uncoated detector measures the background caused by fast neutron leakage from the source. These results show that boron coated CR-39 detectors cannot be used alone for accurate slow neutron measurements. Reliable neutron fluence determination requires the simultaneous use of an uncoated detector. The difference between the two detectors provides a correct estimate of the thermal neutron flux in mixed radiation fields and where conventional neutron detectors cannot be used.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑