可配置的γ光子谱仪用于实现精确的放射引导肿瘤切除
Configurable γ Photon Spectrometer to Enable Precision Radioguided Tumor Resection
AI总结:
本文提出了一种基于集成电路的高灵敏度γ光子谱仪,用于提高放射引导手术中肿瘤切除的精度和准确性。
摘要:
外科肿瘤切除旨在去除肿瘤边缘的所有癌细胞,并在组织表面下厘米级深度内完全清除。在手术过程中,微小的疾病簇难以在术中可视化,往往被遗漏,显著增加了癌症复发的风险。放射引导手术(RGS)已显示出能够使用伽马(γ)光子发射的放射性同位素来标记癌细胞以识别它们的能力,但需要毫米级的γ光子谱仪来在组织边缘(即伽马光子入射能量的函数)中以高特异性定位这些细胞的位置。本文提出了一种基于9.9 mm²集成电路(IC)的γ谱仪,采用180 nm CMOS工艺,用于测量单个γ光子及其入射能量,具有亚千电子伏特的能量分辨率。我们使用小尺寸22微米的反向偏置二极管,这些二极管具有低耗尽区域电容,因此能产生毫伏级的电压信号以响应由入射γ光子产生的小电荷。通过测量生成的电压信号在γ探测事件后恢复到直流电平所需的时间来实现低功耗的能量谱仪方法,而不是直接测量电压下降。该谱仪方法在三种不同的像素架构中实现,允许根据RGS中广泛异质的手术和患者状况配置像素灵敏度、能量分辨率和能量动态范围。该谱仪已使用三种常见的γ发射放射性同位素(64Cu、133Ba、177Lu)进行了测试,并能以亚千电子伏特的能量分辨率和1.315 MeV的能量动态范围,使用5分钟的采集时间分辨出至1微居里的活动。
英文摘要:
Surgical tumor resection aims to remove all cancer cells in the tumor margin and at centimeter-scale depths below the tissue surface. During surgery, microscopic clusters of disease are intraoperatively difficult to visualize and are often left behind, significantly increasing the risk of cancer recurrence. Radioguided surgery (RGS) has shown the ability to selectively tag cancer cells with gamma (γ) photon emitting radioisotopes to identify them, but require a mm-scale γ photon spectrometer to localize the position of these cells in the tissue margin (i.e., a function of incident γ photon energy) with high specificity. Here we present a 9.9 mm2 integrated circuit (IC)-based γ spectrometer implemented in 180 nm CMOS, to enable the measurement of single γ photons and their incident energy with sub-keV energy resolution. We use small 2 2 um reverse-biased diodes that have low depletion region capacitance, and therefore produce millivolt-scale voltage signals in response to the small charge generated by incident γ photons. A low-power energy spectrometry method is implemented by measuring the decay time it takes for the generated voltage signal to settle back to DC after a γ detection event, instead of measuring the voltage drop directly. This spectrometry method is implemented in three different pixel architectures that allow for configurable pixel sensitivity, energy-resolution, and energy dynamic range based on the widely heterogenous surgical and patient presentation in RGS. The spectrometer was tested with three common γ-emitting radioisotopes (64Cu, 133Ba, 177Lu), and is able to resolve activities down to 1 uCi with sub-keV energy resolution and 1.315 MeV energy dynamic range, using 5-minute acquisitions.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。