光谱微CT用于纤维软骨钙化的定量分析

Spectral micro-CT for quantitative analysis of calcification in fibrocartilage

INFN - Division of Ferrara(意大利国家核物理研究所费拉拉分部);Dept. of Physics and Earth Science, University of Ferrara(费拉拉大学地球科学与物理学系);Dept. of Physics, University of Trieste(的里雅斯特大学物理系);INFN - Division of Trieste(意大利国家核物理研究所的里雅斯特分部);RAMSES Laboratory, RIT Department, IRCCS Istituto Ortopedico Rizzoli(IRCCS里佐利骨科医院RIT部门RAMSES实验室);nd Orthopaedic and Traumatologic Clinic, IRCCS Istituto Ortopedico Rizzoli(IRCCS里佐利骨科医院第二骨科与创伤科诊所)

Vittoria Mazzini, Paolo Cardarelli, Andrew L. Coathup, Eleonora Olivotto, Francesco Grassi, Enrico Tassinari, Simone Velardita, Angelo Taibi, Luca Brombal

AI总结:

本研究提出了一种基于光谱微CT的定量方法,用于非破坏性地分析纤维软骨钙化,提供三维钙分布的精确量化。

摘要:

本工作介绍了一种用于评估纤维软骨钙化的定量方法,使用光谱微CT(μCT)。髋关节髋臼杯唇组织样本来自骨关节炎和髋骨髋臼撞击症患者,使用配备小像素光电计数探测器的实验室级光谱μCT系统进行成像。探测器使用两个能量阈值,允许在不同X射线能量下同时获取两个CT数据集。一种考虑系统光谱响应的材料分解算法用于分离钙和水样成分,从而获得完整石蜡嵌片样本中钙化区域的三维可视化和定量分析。与基于组织学的传统钙化评估方法不同,这种光谱μCT方法可以在不进行物理切片或染色的情况下对钙结构进行体积定量。该方法在3厘米大小的样本上实现了20微米的像素尺寸,钙检测阈值为约0.3克/立方厘米,适用于50微米大小的结构。通过使用校准幻影估计了定量的准确性,约为5%。进一步与组织学的比较显示,该方法正确定位了钙的空间分布,并通过提供未破坏组织样本的非破坏性、定量3D钙图,实现了钙晶体沉积评分的匹配。该技术补充了组织学,可以增强髋关节疾病病理纤维软骨钙化的表征。

英文摘要:

This work introduces a quantitative method for assessing calcification in fibrocartilage using spectral micro-computed tomography ($μ$CT). Tissue samples of hip acetabular labrum from patients with osteoarthritis and femoroacetabular impingement were imaged with a laboratory-based spectral $μ$CT system equipped with a small-pixel photon-counting detector. The detector operated with two energy thresholds, allowing the simultaneous acquisition of two CT datasets at different X-ray energies. A material decomposition algorithm accounting for the system's spectral response was applied to separate calcium- and water-like components, yielding three-dimensional visualization and quantification of calcified regions within intact paraffin-embedded samples. Unlike the conventional method for calcification assessment based on histology, this spectral μCT approach offers volumetric quantification of calcium structures without physical sectioning or staining. The method achieved a voxel size of 20 $μ$m for samples up to ~3 cm, with a calcium detection threshold of ~0.3 g/cm$^3$ for structures down to 50 $μ$m. Quantification accuracy was estimated to be 5% by using a calibration phantom. Further comparison with histology demonstrated the correct localization of calcium spatial distributions and a match in the calcium crystal deposition score by providing non-destructive, quantitative 3D calcium maps of preserved tissue samples. This technique complements histology and could enhance the characterization of pathological fibrocartilage calcification in hip joint disorders.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑