暴露于高压后的MKS 390微型电离规性能研究

MKS 390 Micro-Ion Gauge Performance After Exposure to High Pressures

Rochester Institute of Technology(罗切斯特理工学院);Tritium Solutions Inc.(氚解决方案公司)

Brandon Massett, Walter Shmayda

AI总结:

本研究对比了曾暴露于10000 Torr高压与未接触过1000 Torr以上压力的两台MKS Granville-Phillips 390微型电离规,发现高压暴露对其读数影响可忽略,且不会破坏泄漏完整性,验证了该规高压暴露后的功能可靠性。

摘要:

在某些应用场景中,压力传感器可能会有意或无意地暴露于高压环境下,这引发了人们对其后续功能可用性的担忧。我们将两台MKS Granville-Phillips 390微型电离规的性能进行了相互对比,其中一台曾暴露于10000 Torr的压力下,另一台从未暴露于1000 Torr以上的压力环境。结果表明,两台规的读数差异处于制造商规定的不确定度范围内,说明高压暴露带来的影响可忽略不计。此外,高压暴露并未损害该电离规的泄漏完整性。

英文摘要:

In certain applications, pressure transducers may be exposed to high pressures either deliberately or accidentally, raising concerns about their functionality afterwards. We compared the performance of two MKS Granville-Phillips 390 Micro-Ion Gauges against each other, one that had been exposed to 10,000 Torr and the other had never been exposed to pressures above 1000 Torr. Our results show that the differences in the readings between the gauges were within the range of uncertainty specified by the manufacturer indicating negligible impact due to the exposure to high pressure. Additionally, the high pressure exposure did not compromise the leak integrity of the gauge.