相位敏感建模改善脂肪-水混合物中的Fat DESPOT多参数弛豫映射
Phase-sensitive modelling improves Fat DESPOT multiparametric relaxation mapping in fat-water mixtures
AI总结:
本研究针对Fat DESPOT多参数弛豫映射技术,通过升级脂肪-水分离初始猜测方法、引入相位敏感建模,提升了脂肪-水混合物中PDFF及$R_1$等参数的估计精度与准确性。
摘要:
目的:改进Fat DESPOT的原始形式,这是一种多参数映射技术,可返回脂肪和水特异性的$R_1$($R_{1f}$、$R_{1w}$)、$R_2^*$以及质子密度脂肪分数(PDFF)估计值,改进方式为升级用于选择初始参数猜测值的脂肪-水分离方法,并引入对水和脂肪信号相位的显式模型敏感性。方法:我们在3 T场强下,使用可变脂肪分数凝胶体模开展体模实验,对比了3点Dixon法和Graph Cut(GC,图割)法为Fat DESPOT提供初始猜测值的效果。同样在体模实验中,我们将原始Fat DESPOT方法与分别对脂肪和水相位建模的幅度方法(Fat DESPOT$_{mϕ}$)、对复数数据建模的方法(Fat DESPOT$_c$)进行了对比。随后将表现最佳的方法应用于一名健康人类受试者的小腿。结果:在体模中,采用3点Dixon法和GC法的Fat DESPOT在参数估计值和精度方面表现相近,不过Dixon法在标称脂肪分数为50%的感兴趣区(ROI)中偏离了整体趋势。此外,Fat DESPOT$_c$与参考PDFF的一致性最佳(平均误差1.5 ± 1.2%),且在所有ROI中,PDFF、$R_{1f}$和$R_{1w}$的组合标准差最低(σ分别为0.13%、0.19 s$^{-1}$、0.0082 s$^{-1}$)。结论:与其他对比方法相比,本方法的$R_{1f}$和$R_{1w}$精度更高、PDFF准确性更好,且回波时间通用性更强,本研究证明了将GC法用于初始猜测值、结合复数拟合的方案在Fat DESPOT多参数成像中的优势。
英文摘要:
Purpose: To improve on the original form of Fat DESPOT, a multiparametric mapping technique that returns the fat- and water-specific estimates of $R_1$ ($R_{1f}$, $R_{1w}$), $R_2^*$ , and proton density fat fraction (PDFF) by upgrading the fat-water separation method used for selection of initial parameter guesses, and by introducing explicit model sensitivity to the phase of the water and fat signals. Methods: We compared the 3-point Dixon and Graph Cut (GC) approaches to initial guesses for Fat DESPOT in phantom experiments at 3 T in a variable fat fraction gel phantom. Also in phantom, we then compared the original Fat DESPOT approach to a magnitude approach modelling the phases of fat and water separately (Fat DESPOT$_{mϕ}$), and an approach that models the complex data (Fat DESPOT$_c$). The best-performing approach was then used in the lower leg of a healthy human participant. Results: In phantoms, Fat DESPOT using the 3-point Dixon and GC performed similarly in parametric estimates and precision, though the Dixon approach deviated from the overall trend in the 50% nominal fat fraction ROI. Furthermore, Fat DESPOT$_c$ showed the best agreement with reference PDFF (average error 1.5 +/- 1.2%) and the lowest combined standard deviation across ROIs, for PDFF, $R_{1f}$, and $R_{1w}$ (σ = 0.13%, 0.19 s$^{-1}$, 0.0082 s$^{-1}$). Conclusion: With a higher precision of $R_{1f}$ and $R_{1w}$ , accuracy of PDFF, and more echo time versatility than other compared approaches, this work demonstrates the advantages of the GC approach for initial guesses paired with complex fitting for Fat DESPOT multiparametric imaging.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。