髋植入物的计算优化研究:基于密度映射的功能梯度生物仿生TPMS基格栅结构

A Computational Optimisation Study of Hip Implant Using Density Mapping Functionally Graded Biomimetic TPMS-based Lattice Structures

University of Galway(高威大学);Griffith University(格里菲斯大学)

Mahtab Vafaeefar, Conall Quinn, Kevin M. Moerman, Ted J. Vaughan

AI总结:

本研究提出一种基于密度映射的功能梯度生物仿生TPMS格栅结构,通过计算优化减少应力屏蔽,改善骨整合与应力传递。

摘要:

本研究通过开发一种功能梯度生物仿生格栅结构,提出了一种髋植入物的计算优化框架,该结构在结构上被优化以限制应力屏蔽。优化技术受骨重塑算法逆过程的启发,通过减少密度和因此的刚度,从而在应变能高于参考水平的区域内实现均匀的应力分布。优化技术的结果提供了一个非均匀的密度分布场,该分布显示在植入物柄的侧面具有较低的密度水平,在柄的中轴线周围具有较高的材料密度。通过将三重周期最小表面格栅结构映射到植入物上,捕捉了优化的材料分布,从而在固态植入物内部形成了多孔格栅表面。通过实现有限元骨重塑算法,并将骨响应与具有完全实心植入物的股骨进行比较,评估了多孔植入物设计的性能,以应力分布和质量变化为评估标准。分析结果表明,骨-植入物界面处的骨形成得到了改善,且植入物向周围骨传递应力的能力得到了增强。

英文摘要:

This study presents a computational optimisation framework of a hip implant through the development of a functionally graded biomimetic lattice structure, whose design was structurally optimised to limit stress shielding. The optimisation technique was inspired by the inverse of a bone remodelling algorithm, promoting an even stress distribution throughout the design region, by reducing the density and consequently the stiffness, in regions where strain energy was higher than the reference level. The result of the optimisation technique provided a non-uniform graded density distribution field that showed lower density level on the sides of the implant stem, and higher material density around the medial axis of the stem. The optimised material distribution was captured using mapping of a triply periodic minimal surface lattice structure on the implant, which resulted in porous lattice surfaces inside the solid implant. The performance of the porous implant design was evaluated through implementation of a finite element bone remodelling algorithm and comparing the bone response with a femur with fully solid implant model, in terms of stress distribution and mass change. The results of the analysis showed improved bone formation on the bone-implant interface, and enhanced stress transmission to the surrounding bone from the implant.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑