不确定度在Optode位置上对自校准或双斜率漫射光测量中的影响

Influence of Uncertainties in Optode Positions on Self-Calibrating or Dual-Slope Diffuse Optical Measurements

Giles Blaney, Angelo Sassaroli, Tapan Das, Sergio Fantini

AI总结:

本文定量分析了optode位置误差对自校准和双斜率漫射光测量光学特性的影响,发现其对绝对散射系数的影响大于吸收系数,并确定了各几何排列下影响最小的位置不确定度方向,以指导光学探头的优化设计。

摘要:

自校准和双斜率测量已用于漫射光学领域,以稳健评估高散射介质和生物组织光学特性的绝对值或时间变化。这些测量采用具有至少两个光源位置和至少两个探测器位置的光学探头。本工作重点定量分析这些光源和探测器位置的误差对光学特性评估的影响。我们考虑了线性、梯形和矩形optode排列,并基于半无限均匀介质的扩散理论进行了理论计算。我们发现,optode位置的不确定度对绝对散射系数测量的影响可能大于对吸收系数的影响。例如,在线性排列中,将每个optode在任何方向上位移1毫米,绝对吸收和散射分别可预期出现4.1%和19%的平均误差。对于吸收变化的测量,optode位置误差的影响通常较小。对于每种几何排列(线性、梯形、矩形),我们确定了每个optode位置不确定度的方向,使其对光学测量的影响最小。这些结果可指导自校准和双斜率测量的光学探头的优化设计。

英文摘要:

Self-calibrating and dual-slope measurements have been used in the field of diffuse optics for robust assessment of absolute values or temporal changes in the optical properties of highly scattering media and biological tissue. These measurements employ optical probes with a minimum of two source positions and a minimum of two detector positions. This work focuses on a quantitative analysis of the impact of errors in these source and detector positions on the assessment of optical properties. We considered linear, trapezoidal, and rectangular optode arrangements and theoretical computations based on diffusion theory for semi-infinite homogeneous media. We found that uncertainties in optodes' positions may have a greater impact on measurements of absolute scattering versus absorption coefficients. For example, 4.1% and 19% average errors in absolute absorption and scattering, respectively, can be expected by displacing every optode in a linear arrangement by 1 mm in any direction. The impact of optode position errors is typically smaller for measurements of absorption changes. For each geometrical arrangement (linear, trapezoid, rectangular), we identify the direction of the position uncertainty for each optode that has minimal impact on the optical measurements. These results can guide the optimal design of optical probes for self-calibrating and dual-slope measurements.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑