直立质子弧中的穿透层在计划质量和射程验证方面解锁优势
Shoot-through layers in upright proton arcs unlock advantages in plan quality and range verification
AI总结:
本研究针对紧凑型直立质子治疗系统束流半影扩大问题,提出结合静态质子弧与穿透层的计划,通过虚拟体模和头颈部病例验证其在计划质量、鲁棒性及射程验证上的优势,并缩短治疗时间。
摘要:
背景:采用紧凑型传输系统的直立质子治疗有望降低治疗成本,但也可能因靠近患者处的能量选择导致束流半影扩大。目的:本研究旨在将直立静态质子弧与额外的穿透(ST)质子层相结合,以锐化束流半影并改善此类系统的计划质量。方法:我们针对虚拟体模检查了多种治疗计划:3束笔形束扫描质子治疗(IMPT)、带/不带穿透层的静态弧(Arc+ST/Arc)以及带/不带准直的计划(+Coll)。在虚拟体模中使用了三个不同靶区,以研究对适形指数(CI)、均匀性指数(HI)、鲁棒性和体模体积平均剂量的影响。体模研究辅以头颈部(H&N)患者病例,采用类似的计划集。估算了所有计划的传输时间。在头颈部病例的模拟场景中,研究了一种确定穿透质子残余射程的射程验证概念。结果:在体模研究中,与带准直的静态弧(Arc+Coll)计划相比,Arc+ST计划显示出更优的CI、HI和靶区鲁棒性。对于不带ST的Arc计划,准直计划的性能优于非准直计划;而对于Arc+ST,准直对CI、HI和鲁棒性影响很小。在头颈部病例中,Arc+ST也显示出类似的改善。这些结果表明,将弧与ST结合时无需孔径,从而大幅缩短治疗时间。射程验证模拟显示,该方法对检测阻止本领比(SPR)误差、摆位误差和患者解剖结构变化敏感。结论:结合质子弧和穿透层可通过改善计划质量增强紧凑型直立质子解决方案,且该方案还适合纳入快速直接的残余射程验证方法。
英文摘要:
Background: Upright proton therapy with compact delivery systems has the potential to reduce costs for treatments but could also lead to broadening of the beam penumbra due to energy selection close to the patient. Purpose: This study aims at combining upright static proton arcs with additional layers of shoot-through (ST) protons to sharpen the beam penumbra and improve plan quality for such systems. Methods: We examined various treatment plans for a virtual phantom: 3-beam IMPT, static arc (Arc) with/without ST (Arc+ST), and with/without collimation (+Coll). In the virtual phantom three different targets were utilized to study the effect on conformity index (CI), homogeneity index (HI), robustness and mean dose to the phantom volume. The phantom study was complemented with a head-and-neck (H&N) patient case with a similar set of plans. The delivery time for all plans was estimated. A range verification concept that determines residual ranges of the ST protons was studied in simulated scenarios for the H&N case. Results: In the phantom study, the Arc+ST plans show superior CI, HI and target robustness compared to the Arc+Coll plans. For the Arc plans without ST, the collimated plans perform better than the uncollimated plans. On the other hand, for Arc+ST, collimation has little impact on CI, HI and robustness. For the H&N case, similar improvements for Arc+ST can be seen. These results imply that no aperture is needed when combining arcs with ST, which in turn substantially reduces treatment times. The range verification simulation shows that the method is sensitive to detect SPR errors, setup errors and changes in the patient anatomy. Conclusions: Combining proton arcs and ST layers can enhance compact upright proton solutions by improving plan quality. It is also tailored for the inclusion of a fast and straightforward residual range verification method.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。