基于拟蒙特卡洛的CT重建散射校正
Scatter correction based on quasi-Monte Carlo for CT reconstruction
AI总结:
本文提出FDK-QMC-BM4D多模块耦合CT重建框架,利用QMC高效计算高阶散射积分并校正能谱,可同时消除散射、射束硬化伪影和噪声,速度较Monte Carlo提升约102倍。
摘要:
散射信号会降低计算机断层扫描(CT)图像的对比度和分辨率,并诱发伪影。如何有效校正CT中的散射信号,一直是研究人员关注的焦点。本工作提出了一种用于消除CT中散射伪影的新框架。在该框架中,光子与物质之间的相互作用被表征为一个Markov过程,而CT中散射信号强度的计算被转化为一个$4n$维积分的计算,其中$n$是最高散射阶数。鉴于CT中散射信号的低频特性,本文采用拟蒙特卡洛(QMC)方法,并结合强制固定探测和下采样来计算该积分。在重建过程中,考虑了散射信号对X射线能谱的影响。提出了一种经散射校正的能谱估计方法,并将其用于估计X射线能谱。基于Feldkamp-Davis-Kress(FDK)算法,开发了一种多模块耦合重建方法,称为FDK-QMC-BM4D,以同时消除CT成像中的散射伪影、射束硬化伪影和噪声。最后,在Shepp-Logan体模和头部中验证了FDK-QMC-BM4D方法的有效性。与目前估计和校正CT散射信号最准确、并被广泛认可的Monte Carlo方法相比,FDK-QMC-BM4D方法在保证精度的同时,将运行速度提高了约$102$倍。通过整合FDK-QMC-BM4D的机制,本研究为解决临床CT中的伪影问题提供了一种新途径。
英文摘要:
Scatter signals can degrade the contrast and resolution of computed tomography (CT) images and induce artifacts. How to effectively correct scatter signals in CT has always been a focal point of research for researchers. This work presents a new framework for eliminating scatter artifacts in CT. In the framework, the interaction between photons and matter is characterized as a Markov process, and the calculation of the scatter signal intensity in CT is transformed into the computation of a $4n$-dimensional integral, where $n$ is the highest scatter order. Given the low-frequency characteristics of scatter signals in CT, this paper uses the quasi-Monte Carlo (QMC) method combined with forced fixed detection and down sampling to compute the integral. In the reconstruction process, the impact of scatter signals on the X-ray energy spectrum is considered. A scatter-corrected spectrum estimation method is proposed and applied to estimate the X-ray energy spectrum. Based on the Feldkamp-Davis-Kress (FDK) algorithm, a multi-module coupled reconstruction method, referred to as FDK-QMC-BM4D, has been developed to simultaneously eliminate scatter artifacts, beam hardening artifacts, and noise in CT imaging. Finally, the effectiveness of the FDK-QMC-BM4D method is validated in the Shepp-Logan phantom and head. Compared to the widely recognized Monte Carlo method, which is the most accurate method by now for estimating and correcting scatter signals in CT, the FDK-QMC-BM4D method improves the running speed by approximately $102$ times while ensuring accuracy. By integrating the mechanism of FDK-QMC-BM4D, this study offers a novel approach to addressing artifacts in clinical CT.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。