磁共振测温中的焦点校正用于聚焦超声丘脑切开术的精准靶向
Focus Correction in MR thermography for Precise Targeting in Focused Ultrasound Thalamotomy
AI总结:
本文提出一种两步校正方法,利用场图和TE误差图补偿化学位移与k空间偏移引起的热点位置偏差,将平均温度误差降至-0.05°C,从而提升聚焦超声丘脑切开术的靶向精度。
摘要:
目的:通过校正靶点坐标与磁共振测温中所见热点之间由温度变化引起的空间偏差,来调整消融焦点的位置。方法:采用两步校正程序来处理化学位移和k空间偏移。第一步利用焦点周围的场图逐像素校正空间偏移。第二步计算温度引起的TE变化图,以补偿焦点中心位置的偏移。结果:由化学位移和k空间中心偏移引起的约1mm偏移通过场图和TE误差图进行了校正。平均温度误差降至-0.05°C,而仅使用化学位移校正时为-0.11°C。结论:校正温度变化中场梯度和k空间偏移引起的热点空间偏移,可以确保聚焦超声(FUS)丘脑切开术中的精准靶向。
英文摘要:
Purpose: To adjust the location of the ablation focus by correcting the spatial discrepancy induced by temperature changes between the target coordinate and the hotspot as seen in MR thermography. Methods: A two-step correction procedure was used to account for the chemical and k-space shifts. The first step corrected the spatial shift pixel by pixel using the field map around the focus. The second step calculated the temperature-induced TE variation map to compensate for the center location offset of the focus. Results: The approximately 1mm shift caused by chemical shift and k-space center offset was adjusted using the field map and TE error map. The mean temperature error was reduced to -0.05°C, as opposed to -0.11°C obtained using only a chemical shift correction. Conclusion: Correcting for the spatial shifts of hotspots caused by field gradients and k-space shifts in temperature changes can ensure precise targeting in FUS thalamotomy.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。