电光调制器作为无样品校准和频率稳定器用于布里渊显微镜

Electro-Optic Modulator source as sample-free calibrator and frequency stabilizer for Brillouin Microscopy

Center for Life Nano- and Neuro-Science, Istituto Italiano di Tecnologia(生命纳米与神经科学中心,意大利理工大学);CrestOptics S.p.A(CrestOptics公司);Stem Cell Medicine Department, Murdoch Children’s Research Institute(儿童健康研究所干细胞医学部);Department of Biology and Biotechnology Charles Darwin, Sapienza University of Rome(罗马大学查尔斯·达尔文生物与生物技术系);Department of Molecular Medicine, Sapienza University of Rome(罗马大学分子医学系);Dipartimento Di Fisica, Sapienza University of Rome(罗马大学物理系)

Claudia Testi, Emanuele Pontecorvo, Chiara Bartoli, Chiara Marzaro, Fabrizio Gala, Li Zhang, Giulia Zanini, Giancarlo Ruocco

AI总结:

本文提出利用电光调制器作为无样品校准和频率稳定器,提升布里渊显微镜的稳定性和精度,实现自动数据采集和时间漂移补偿。

摘要:

布里渊显微镜是一种新型无标记光学技术,能够非侵入性地测量材料在亚微米尺度上的机械性能;近年来,其在生命科学中的应用已大幅扩展。目前,许多定制的布里渊显微镜存在时间不稳定性,影响性能,导致获取的光谱在时间上出现漂移,可能使不同日期的数据不一致。标准布里渊显微镜的校准问题在于,文献中常用的协议使用已知的布里渊位移(水和甲醇)来重建色散曲线,但这种方法受外部因素影响较大,与光谱仪性能无关。因此需要手动和频繁地重新对齐光谱仪并用标准材料进行重复校准。本文展示了一种创新方法,通过插入电光调制器(EOM)来消除标准布里渊显微镜的时间不稳定性:i)在测量过程中用作参考信号;ii)作为校准器,允许在自动流程中高精度重建光谱仪色散曲线,无需参考样品;iii)作为检测和补偿时间漂移的工具,通过闭环反馈控制。我们展示,配备EOM和可调激光器的布里渊显微镜能够自动采集超过2天的数据,无需重新对齐光谱仪;获取的布里渊位移和宽度显示出比标准布里渊显微镜更优异的时间稳定性。

英文摘要:

Brillouin Microscopy is a novel label-free optical technique that enables the measurement of a material's mechanical properties at the sub-micron scale in a non-invasive and non-contact way; in the last few years, its applications in the life sciences have extensively expanded. To date, many custom-built Brillouin Microscopes suffer from temporal instabilities that impact their performances, showing drifts in the acquired spectra during time that may lead to inconsistencies between data acquired at different days. A further challenge for standard Brillouin Microscopes is the calibration of the spectrometer: the currently accepted protocol in literature uses known Brillouin shifts of water and methanol to reconstruct the dispersion curve, but this approach is highly influenced by external factors that are unrelated to spectrometer's performances. Manual and frequent realignments of the spectrometer and repeated calibrations with standard materials are thus needed to address these issues. Here, we show an innovative method to remove temporal instabilities of a standard Brillouin Microscope by inserting an Electro-Optic Modulator (EOM) that can be used: i) as a reference signal during measurements; ii) as a calibrator, allowing the reconstruction of the spectrometer dispersion curve with high precision, in an automatic pipeline and without the need for reference samples; iii) as a tool to detect and compensate for temporal drifts through a feedback control in a closed loop. We here show that our Brillouin Microscope, equipped with an EOM and a tuneable laser, is able to automatically acquire data for more than 2 days without the need to realign the spectrometer; retrieved Brillouin shifts and widths showed superior stability in time than standard Brillouin Microscopes.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑