arXivDaily arXiv每日学术速递 周一至周五更新
arXiv 2411.04515physics.med-phcs.SYeess.SY
收藏

对接式种植体支持修复体几何形状对修复螺钉疲劳寿命的影响

Effect of the geometry of butt-joint implant-supported restorations on the fatigue life of prosthetic screws

Mikel Armentia, Mikel Abasolo, Ibai Coria, Nicolas Saintier

AI总结:

本研究通过疲劳测试和线性回归分析,评估了种植体几何参数(体部直径、平台直径、连接类型)对修复螺钉疲劳寿命的影响,发现增大直径可显著提升疲劳寿命,且内外连接效果相当。

摘要:

问题陈述。种植体几何形状影响对接式种植体支持修复体的力学性能和疲劳行为。然而,种植体部件的失效通常是在忽略修复螺钉的情况下进行研究的,而修复螺钉往往是关键的修复部件。目的。评估3个主要种植体几何参数的影响:种植体体部直径、平台直径和种植体-基台连接类型(外部对接与内部对接)对修复螺钉疲劳寿命的影响。将实验值与使用先前发表的方法论获得的理论值进行比较。材料与方法。测试了制造商BTI的4种不同的直接种植体牙科修复设计。根据ISO 14801标准,在轴向疲劳试验机上进行了48次疲劳测试。获得了线性回归模型、线性回归的95%置信区间带以及疲劳载荷-寿命结果的95%预测区间,并通过协方差分析进行比较,以确定所研究的3个参数对疲劳行为的影响。结果。线性回归模型显示,当种植体体部直径增加1毫米时,存在统计学差异;观察到疲劳寿命平均增加3.5倍。将种植体基台连接直径增加1.4毫米也显示出显著差异,导致疲劳寿命平均延长7倍。没有发现显著的统计学证据表明内部和外部连接之间的疲劳寿命存在差异。结论。增加种植体平台和体部直径显著改善了螺钉的疲劳寿命,而外部和内部连接提供了相似的结果。此外,实验结果证明了疲劳寿命预测方法的准确性。

英文摘要:

Statement of problem. Dental implant geometry affects the mechanical performance and fatigue behavior of butt-joint implant-supported restorations. However, failure of the implant component has been generally studied by ignoring the prosthetic screw, which is frequently the critical restoration component Purpose. Evaluate the effect of 3 main implant geometric parameters: the implant body diameter, the platform diameter, and the implant-abutment connection type (external versus internal butt-joint) on the fatigue life of the prosthetic screw. The experimental values were further compared with the theoretical ones obtained by using a previously published methodology M&M. 4 different designs of direct-to-implant dental restorations from the manufacturer BTI were tested. Forty-eight fatigue tests were performed in an axial fatigue testing machine according to ISO 14801. Linear regression models, 95% interval confidence bands for the linear regression, and 95% prediction intervals of the fatigue load-life results were obtained and compared through an analysis of covariance to determine the influence of the 3 parameters under study on the fatigue behavior Results. Linear regression models showed a statistical difference when the implant body diameter was increased by 1 mm; an average 3.5-fold increase in fatigue life was observed. Increasing the implant abutment connection diameter by 1.4 mm also showed a significant difference, leading to 7-fold longer fatigue life on average. No significant statistical evidence was found to demonstrate a difference in fatigue life between internal and external connections Conclusions. Increasing the implant platform and body diameter significantly improved the fatigue life of the screw, whereas external and internal connections provided similar results. In addition, experimental results proved the accuracy of the fatigue life prediction methodology

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑