确定可穿戴脑磁图系统中的传感器几何结构与增益

Determining sensor geometry and gain in a wearable MEG system

Ryan M. Hill, Gonzalo Reina Rivero, Ashley J. Tyler, Holly Schofield, Cody Doyle, James Osborne, David Bobela, Lukas Rier, Joseph Gibson, Zoe Tanner, Elena Boto… 展开作者

Ryan M. Hill, Gonzalo Reina Rivero, Ashley J. Tyler, Holly Schofield, Cody Doyle, James Osborne, David Bobela, Lukas Rier, Joseph Gibson, Zoe Tanner, Elena Boto, Richard Bowtell, Matthew J. Brookes, Vishal Shah, Niall Holmes

AI总结:

针对可穿戴脑磁图系统中OPM传感器位置与增益校准的复杂性,提出HALO和矩阵线圈两种磁场生成校准方法,实验证明两者均能准确校准传感器阵列并显著提升信噪比,推动OPM-MEG向常规应用迈进。

摘要:

光泵磁力计(OPMs)是紧凑轻便的传感器,能够测量大脑神经元集合中电流流动产生的磁场。此类传感器使得脑磁图(MEG)仪器的构建成为可能,相比传统MEG设备具有显著优势,包括对头围的适应性、增强的运动容忍度、更低的复杂性以及改善的数据质量。然而,实现OPMs的潜力取决于我们进行系统校准的能力,即确定传感器位置、方向以及传感器输出与磁场之间的关系(称为传感器增益)。这种校准在OPM-MEG中较为复杂,例如,OPM的放置会因受试者而异(不同于传统MEG中传感器位置或方向是固定的)。在此,我们提出了两种校准方法,均基于在传感器阵列中生成特性明确的磁场。我们的第一种设备(HALO)是一种头戴式系统,通过一组线圈生成类偶极子磁场。第二种设备(矩阵线圈(MC))利用嵌入在磁屏蔽室墙壁中的线圈生成磁场。我们的结果表明,两种方法均提供了校准OPM阵列的准确手段(例如,传感器位置在2毫米的真实值范围内),且两种方法生成的校准结果彼此高度一致。当应用于人类MEG实验数据时,两种方法在波束成形后均提供了改善的信噪比,表明它们提供的校准参数比出厂设置和假定的物理传感器坐标与方向更接近真实值。这两种技术均实用且易于集成到现实世界的MEG应用中。这显著推进了OPMs在MEG记录中的常规应用进程。

英文摘要:

Optically pumped magnetometers (OPMs) are compact and lightweight sensors that can measure magnetic fields generated by current flow in neuronal assemblies in the brain. Such sensors enable construction of magnetoencephalography (MEG) instrumentation, with significant advantages over conventional MEG devices including adaptability to head size, enhanced movement tolerance, lower complexity and improved data quality. However, realising the potential of OPMs depends on our ability to perform system calibration, which means finding sensor locations, orientations, and the relationship between the sensor output and magnetic field (termed sensor gain). Such calibration is complex in OPMMEG since, for example, OPM placement can change from subject to subject (unlike in conventional MEG where sensor locations or orientations are fixed). Here, we present two methods for calibration, both based on generating well-characterised magnetic fields across a sensor array. Our first device (the HALO) is a head mounted system that generates dipole like fields from a set of coils. Our second (the matrix coil (MC)) generates fields using coils embedded in the walls of a magnetically shielded room. Our results show that both methods offer an accurate means to calibrate an OPM array (e.g. sensor locations within 2 mm of the ground truth) and that the calibrations produced by the two methods agree strongly with each other. When applied to data from human MEG experiments, both methods offer improved signal to noise ratio after beamforming suggesting that they give calibration parameters closer to the ground truth than factory settings and presumed physical sensor coordinates and orientations. Both techniques are practical and easy to integrate into real world MEG applications. This advances the field significantly closer to the routine use of OPMs for MEG recording.