心脏CT图像中的小型金属伪影检测与修复
Small metal artifact detection and inpainting in cardiac CT images
AI总结:
针对植入式除颤器金属导线伪影影响心脏CT量化的问题,提出基于2D U-Net和3D图像修复的深度学习方法,实现伪影自动检测与解剖结构修复,显著提升了心腔分割精度。
摘要:
背景:对于接受立体定向心律失常放射治疗的患者,由于植入式心律转复除颤器(ICDs)金属导线产生的图像伪影,基于治疗前CT图像量化心脏运动十分困难。需要新方法准确减少已重建CT中的金属伪影,以恢复原本丢失的解剖信息。目的:开发一种方法,自动检测心脏CT扫描中的金属伪影,并用解剖学一致的结构和数值修复受影响的区域。方法:收集了12名因治疗室性心动过速而接受心脏放射治疗患者的ECG门控4DCT扫描图像。图像中的金属伪影被手动勾画。开发了一个2D U-Net深度学习(DL)模型来分割金属伪影。通过将患者图像中的金属伪影添加到无伪影的CT中,制备了合成CT数据集。训练了一个3D图像修复DL模型,用真实数值重新填充合成图像中的金属伪影部分。通过分析合成数据集上四个心腔的自动分割结果,对修复模型进行了评估。此外,还对原始心脏患者病例进行了定性检查。结果:伪影检测模型的Dice得分为0.958 ± 0.008。修复模型能够以0.988 ± 0.012的结构相似性指数重建图像。通过心腔分割,表面Dice得分从0.684 ± 0.247提高到0.964 ± 0.067,Hausdorff距离从3.4 ± 3.9 mm降至0.7 ± 0.7 mm。对修复模型在心脏患者CT上的应用进行了视觉检查,伪影修复后的图像在视觉上是合理的。结论:我们成功开发了两个深度模型,用于检测和修复心脏CT图像中的金属伪影。
英文摘要:
Background: Quantification of cardiac motion on pre-treatment CT imaging for stereotactic arrhythmia radiotherapy patients is difficult due to the presence of image artifacts caused by metal leads of implantable cardioverter-defibrillators (ICDs). New methods are needed to accurately reduce the metal artifacts in already reconstructed CTs to recover the otherwise lost anatomical information. Purpose: To develop a methodology to automatically detect metal artifacts in cardiac CT scans and inpaint the affected volume with anatomically consistent structures and values. Methods: ECG-gated 4DCT scans of 12 patients who underwent cardiac radiation therapy for treating ventricular tachycardia were collected. The metal artifacts in the images were manually contoured. A 2D U-Net deep learning (DL) model was developed to segment the metal artifacts. A dataset of synthetic CTs was prepared by adding metal artifacts from the patient images to artifact-free CTs. A 3D image inpainting DL model was trained to refill the metal artifact portion in the synthetic images with realistic values. The inpainting model was evaluated by analyzing the automated segmentation results of the four heart chambers on the synthetic dataset. Additionally, the raw cardiac patient cases were qualitatively inspected. Results: The artifact detection model produced a Dice score of 0.958 +- 0.008. The inpainting model was able to recreate images with a structural similarity index of 0.988 +- 0.012. With the chamber segmentations improved surface Dice scores from 0.684 +- 0.247 to 0.964 +- 0.067 and the Hausdorff distance reduced from 3.4 +- 3.9 mm to 0.7 +- 0.7 mm. The inpainting model's use on cardiac patient CTs was visually inspected and the artifact-inpainted images were visually plausible. Conclusion: We successfully developed two deep models to detect and inpaint metal artifacts in cardiac CT images.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。