混合现实远程超声检查跨越750公里:可行性研究

Mixed Reality Tele-Ultrasound over 750 km: A Feasibility Study

School of Biomedical Engineering, The University of British Columbia(生物医学工程学院,不列颠哥伦比亚大学);Department of Electrical and Computer Engineering, The University of British Columbia(电气与计算机工程系,不列颠哥伦比亚大学);The University of British Columbia(不列颠哥伦比亚大学);Department of Radiology, The University of British Columbia(放射学系,不列颠哥伦比亚大学)

Ryan Yeung, David Black, Patrick B. Chen, Victoria Lessoway, Janice Reid, Sergio Rangel-Suarez, Silvia D. Chang, Septimiu E. Salcudean

AI总结:

本研究提出并评估了一种基于混合现实和触觉反馈的人机远程超声系统,通过新手操作员在专家远程控制下完成腹部超声检查,在754公里距离上实现了92%的图像质量达标率。

摘要:

为解决偏远社区缺乏超声检查的问题,先前工作引入了人机远程操作,一种基于混合现实和触觉的远程超声系统。该方法中,新手扮演认知机器人角色,由专家通过混合现实远程控制。本文总结了该系统的新进展,并描述了一项评估其用于长距离远程腹部超声检查的可行性研究。为提供简单有效的触觉反馈,我们使用了患者椭球模型,并通过系统的位置和力传感器校准其参数。我们在加拿大海达瓜依的斯基德盖特测试了该系统,专家位于754公里外的加拿大温哥华。我们共进行了11次扫描,涉及10名新手和2名超声技师。超声技师的任务是获取上腹部区域的5个目标图像。图像采集质量由2名放射科医生评估。我们收集了对准数据,新手完成了任务负荷和可用性问卷。新手和超声技师均提供了书面和口头反馈,以指导未来的设计迭代。92%的获取图像具有足够质量,可供两位放射科医生解读。新手报告的平均任务负荷低于文献中的参考值,可用性一致获得正面评价。未发现图像质量与跟随者相对于虚拟换能器的对准误差之间存在相关性。总体而言,我们表明人机远程操作使超声技师能够以高性能执行远程腹部超声成像,即使跨越远距离且使用新手跟随者。未来工作将把人机远程操作与传统、机器人及远程指导超声进行比较。

英文摘要:

To address the lack of access to ultrasound in remote communities, previous work introduced human teleoperation, a mixed reality and haptics-based tele-ultrasound system. In this approach, a novice takes the role of a cognitive robot controlled remotely by an expert through mixed reality. In this manuscript we summarize new developments to this system and describe a feasibility study assessing its use for long-distance remote abdominal ultrasound examinations. To provide simple but effective haptic feedback, we used an ellipsoid model of the patient with its parameters calibrated using our system's position and force sensors. We tested the system in Skidegate, Haida Gwaii, Canada, with the experts positioned 754 km away in Vancouver, Canada. We performed 11 total scans with 10 novices and 2 sonographers. The sonographers were tasked with acquiring 5 target images in the epigastric region. The image acquisition quality was assessed by 2 radiologists. We collected alignment data and the novices completed task load and usability questionnaires. Both the novices and sonographers provided written and verbal feedback to inform future design iterations. 92% of the acquired images had sufficient quality for interpretation by both radiologists. The mean task load reported by the novices was below reference values reported in literature and the usability was unanimously positive. No correlation was found between image quality and the follower's alignment error with the virtual transducer. Overall, we show that human teleoperation enables sonographers to perform remote abdominal ultrasound imaging with high performance, even across large distances and with novice followers. Future work will compare human teleoperation to conventional, robotic and tele-mentored ultrasound.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑