用于精密LCR表校准的、在高达2 MHz频率下具有可计算且几乎可忽略的交直流差的电阻标准
Resistance standards with calculable, nearly negligible AC-DC difference at frequencies up to 2 MHz for the calibration of precision LCR meters
AI总结:
本研究开发了基于薄膜SMD电阻串联的新型电阻标准,在2 MHz以下交直流差仅百万分之几,较传统器件精度高四个数量级,可实现精密LCR表的高精度校准,支撑多项阻抗计量新应用。
摘要:
我们开发了基于薄膜表面贴装器件(SMD)电阻器的新型阻抗标准。由于此类电阻器尺寸小巧,决定其频率特性的物理量非常小,既可以通过实验测量,也可以通过数值计算得出。将薄膜SMD电阻器串联使用,能够进一步改善其直流与交流性能。我们本次应用的标称电阻值为12.906千欧,不过其他标称值也同样可以实现。在高达2 MHz的频率范围内,计算得到的频率特性偏差仅为直流阻值的百万分之几,比所有具有相近标称直流阻值的传统可计算交直流电阻器小约四个数量级。为测量频率特性,我们使用了一台工作频率最高达2 MHz的精密电感-电容-电阻(LCR)表,该仪器的测量重复性为百万分之几,但其系统不确定度根据厂商规格,从低频段的300 ppm上升至2 MHz时的3000 ppm。通过对两款结构差异极大的基于SMD的电阻标准进行测量,我们既验证了模型计算结果,也完成了对精密LCR表的校准,二者在整个频率范围内的相对不确定度均为百万分之几。这种精度的提升为该频段开辟了新的应用场景,例如传统可计算电阻标准的验证、阻抗标准的校准,以及未来的量子化霍尔电阻测量。
英文摘要:
We have developed novel impedance standards based on thin-film surface-mount-device (SMD) resistors. Due to the small dimensions of such resistors, the quantities determining their frequency dependence are very small and can be either measured or numerically calculated. A series connection of thin-film SMD resistors allows us to further improve the DC and the AC properties. The nominal resistance value of our application is 12.906 kohm but other values are just as possible. At frequencies up to 2 MHz, the calculated frequency dependence amounts to only a few parts per million of the DC value, which is about four orders of magnitude smaller than for all conventional calculable AC-DC resistors having a similar nominal DC value. To measure the frequency dependence, we use a precision inductance-capacitance-resistance (LCR) meter at frequencies up to 2 MHz that has a reproducibility of a few parts per million but a systematic uncertainty which is specified by the manufacturer to increase from 300 parts per million in the lower frequency range to 3000 parts per million at 2 MHz. Measurements of two very different SMD-based resistance standards allow verification of the model calculation as well as the calibration of the precision LCR meter, both with a relative uncertainty of a few parts per million in the whole frequency range. This boost in precision enables new applications in this frequency range such as the verification of conventional calculable resistance standards, the calibration of impedance standards, and future measurements of the quantum Hall resistance.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。