利用弹性流动不稳定性提升色谱柱动力学性能
Harnessing an elastic flow instability to improve the kinetic performance of chromatographic columns
AI总结:
本研究通过向HPLC柱流动相中添加半稀量HPAM聚合物产生弹性流动不稳定性,证实其可降低峰不对称性、提升柱效率,为开发提升HPLC分离度的聚合物方案奠定基础。
摘要:
尽管经过数十年的研究与开发,淤浆填充的高效液相色谱(HPLC)柱的最优效率仍受到柱壁到中心主体区域固有的长程流动不均匀性的限制。本文展示了一种解决该问题的方案:在装填有10 μm BEH™ 125 Å颗粒的4.6 mm × 300 mm色谱柱运行过程中,向流动相(1% NaCl水溶液)中直接添加半稀剂量(500 ppm)的大尺寸柔性合成聚合物(18 MDa部分水解聚丙烯酰胺,HPAM)。聚合物的加入赋予流动相弹性,使得颗粒间孔隙空间中的流动在超过阈值流速时变得不稳定。我们通过摩擦因子随雷诺数变化的压降测量,验证了这种弹性流动不稳定性的发展。在之前的工作中,我们发现这种流动不稳定性的特征是孔隙尺度流速存在大幅的时空波动,这可能促进分析物在色谱柱内的扩散。对准非保留示踪剂硫脲的轴向扩散测量证实了这一可能性:测量结果明确显示,在不稳定性起始点以上运行可显著降低峰不对称性,从而提升色谱柱效率。这些实验由此提供了概念验证,表明弹性流动不稳定性可被用于减轻跨柱流动不均匀性对淤浆填充HPLC柱效率的负面影响。尽管该方法存在其固有的局限性和约束条件,但本研究为未来定向开发可在常用液相色谱流动相中溶解并赋予流动相弹性的聚合物奠定了基础,这类聚合物可产生弹性流动不稳定性,从而帮助提升HPLC柱的分离度。
英文摘要:
Despite decades of research and development, the optimal efficiency of slurry-packed HPLC columns is still hindered by inherent long-range flow heterogeneity from the wall to the central bulk region of these columns. Here, we show an example of how this issue can be addressed through the straightforward addition of a semidilute amount (500~ppm) of a large, flexible, synthetic polymer (18~MDa partially hydrolyzed polyacrylamide, HPAM) to the mobile phase (1\% NaCl aqueous solution) during operation of a 4.6 mm $\times$ 300 mm column packed with 10~$μ$m BEH$^{\mathrm{TM}}$ 125~Å\ Particles. Addition of the polymer imparts elasticity to the mobile phase, causing the flow in the interparticle pore space to become unstable above a threshold flow rate. We verify the development of this elastic flow instability using pressure drop measurements of the friction factor versus Reynolds number. In prior work, we showed that this flow instability is characterized by large spatiotemporal fluctuations in the pore-scale flow velocities that may promote analyte dispersion across the column. Axial dispersion measurements of the quasi non-retained tracer thiourea confirm this possibility: they unequivocally reveal that operating above the onset of the instability improves column efficiency by significantly reducing peak asymmetry. These experiments thereby provide a proof-of-concept demonstration that elastic flow instabilities can be harnessed to mitigate the negative impact of trans-column flow heterogeneities on the efficiency of slurry-packed HPLC columns. While this approach has its own inherent limitations and constraints, our work lays the groundwork for future targeted development of polymers that can impart elasticity when dissolved in commonly used liquid chromatography mobile phases, and can thereby generate elastic flow instabilities to help improve the resolution of HPLC columns.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。