窃取之术:将为一台超声扫描仪开发的深度学习模型盗取至竞争对手机器
The Art of the Steal: Purloining Deep Learning Models Developed for an Ultrasound Scanner to a Competitor Machine
AI总结:
本文提出一种黑盒无监督域自适应方法,仅通过输入-输出接口将定量超声深度学习模型功能从一台机器迁移至另一台机器,并在实验中实现98%分类准确率,揭示了临床部署中的安全风险。
摘要:
传递函数方法最近已被证明在定量超声(QUS)中校准深度学习(DL)算法方面是有效的,能够解决采集层面和机器层面的数据偏移问题。在此方法的基础上,我们开发了一种策略,在定量超声背景下,将一台超声机器上的深度学习模型功能“窃取”并实施到另一台机器上。这表明深度学习模型的功能可以很容易地在机器之间转移,凸显了将此类模型部署到用于临床用途的商业扫描仪中所带来的安全风险。所提出的方法是一种黑盒无监督域自适应技术,它将传递函数方法与迭代模式相结合。该方法不利用任何与受害机器模型内部相关的信息,而仅依赖于输入-输出接口的可用性。此外,我们假设来自测试机器(即实施窃取的机器)的无标签数据是可用的。随着各公司开始将其深度学习功能部署用于临床用途,这种情况可能变得普遍。竞争公司可能获取受害机器,并通过输入-输出接口将其功能复制到自己的机器上。在实验中,我们使用了一台 SonixOne 和一台 Verasonics 机器。受害模型在 SonixOne 数据上训练,然后其功能被转移到 Verasonics 机器上。所提出的方法成功地将功能转移到了 Verasonics 机器上,在二分类决策任务中取得了显著的 98% 分类准确率。本研究强调,在临床环境中部署深度学习模型之前,有必要建立安全措施。
英文摘要:
A transfer function approach has recently proven effective for calibrating deep learning (DL) algorithms in quantitative ultrasound (QUS), addressing data shifts at both the acquisition and machine levels. Expanding on this approach, we develop a strategy to 'steal' the functionality of a DL model from one ultrasound machine and implement it on another, in the context of QUS. This demonstrates the ease with which the functionality of a DL model can be transferred between machines, highlighting the security risks associated with deploying such models in a commercial scanner for clinical use. The proposed method is a black-box unsupervised domain adaptation technique that integrates the transfer function approach with an iterative schema. It does not utilize any information related to model internals of the victim machine but it solely relies on the availability of input-output interface. Additionally, we assume the availability of unlabelled data from the testing machine, i.e., the perpetrator machine. This scenario could become commonplace as companies begin deploying their DL functionalities for clinical use. Competing companies might acquire the victim machine and, through the input-output interface, replicate the functionality onto their own machines. In the experiments, we used a SonixOne and a Verasonics machine. The victim model was trained on SonixOne data, and its functionality was then transferred to the Verasonics machine. The proposed method successfully transferred the functionality to the Verasonics machine, achieving a remarkable 98\% classification accuracy in a binary decision task. This study underscores the need to establish security measures prior to deploying DL models in clinical settings.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。