University of California, Los Angeles(加州大学洛杉矶分校);Texas A&M University(得克萨斯农工大学);Florida International University(佛罗里达国际大学);University of Connecticut(康涅狄格大学)
Artem Goncharov, Zoltan Gorocs, Ridhi Pradhan, Brian Ko, Ajmal Ajmal, Andres Rodriguez, David Baum, Marcell Veszpremi, Xilin Yang, Maxime Pindrys, Tianle Zheng,… 展开作者

Artem Goncharov, Zoltan Gorocs, Ridhi Pradhan, Brian Ko, Ajmal Ajmal, Andres Rodriguez, David Baum, Marcell Veszpremi, Xilin Yang, Maxime Pindrys, Tianle Zheng, Oliver Wang, Jessica C. Ramella-Roman, Michael J. McShane, Aydogan Ozcan

AI总结:

本研究开发了集成可植入磷光生物传感器与定制紧凑型低成本磷光寿命成像仪的计算型连续葡萄糖监测系统,结合神经网络模型实现耐受错位的血糖推断,体外测试准确率达88.8%,有望成为可穿戴诊断工具。

摘要:

与传统电化学式连续葡萄糖监测(CGM)系统相比,光学CGM系统因成本更低、耐用性更强,正逐步应用于个性化血糖管理。本文提出了一种计算型CGM系统,该系统集成了基于生物相容性磷光材料的可植入式生物传感器,以及一款定制设计的磷光寿命成像仪(PLI)。这款紧凑型低成本PLI的设计目标是透过皮肤采集可植入传感器的磷光寿命图像,其中发射的磷光信号寿命会受局部葡萄糖浓度调控。由于该磷光信号的寿命远长于组织自发荧光或激发光泄漏过程的持续时间,因此可在激发光关闭后的数十微秒内采集传感器发射信号,从而完全规避这些噪声源。透过皮肤采集的寿命图像会经基于神经网络的模型处理,以实现耐受错位的葡萄糖浓度推断,可准确识别正常、低(低血糖)和高(高血糖)浓度范围。我们采用模拟人体皮肤光学特性的1毫米厚皮肤仿体,对添加葡萄糖的样本开展了PLI体外测试,推断准确率达88.8%;同时验证了该系统在皮肤仿体下方可植入传感器位置周围约4.7毫米横向距离内,对随机及未知错位具有鲁棒性。此外,PLI可准确识别超过5毫米的较大横向错位,提示用户介入进行重新校准。这款紧凑型低成本磷光寿命成像仪具备耐受错位的葡萄糖浓度推断能力,是用于实时监测葡萄糖及其他生物标志物的极具吸引力的可穿戴诊断工具。

英文摘要:

Optical continuous glucose monitoring (CGM) systems are emerging for personalized glucose management owing to their lower cost and prolonged durability compared to conventional electrochemical CGMs. Here, we report a computational CGM system, which integrates a biocompatible phosphorescence-based insertable biosensor and a custom-designed phosphorescence lifetime imager (PLI). This compact and cost-effective PLI is designed to capture phosphorescence lifetime images of an insertable sensor through the skin, where the lifetime of the emitted phosphorescence signal is modulated by the local concentration of glucose. Because this phosphorescence signal has a very long lifetime compared to tissue autofluorescence or excitation leakage processes, it completely bypasses these noise sources by measuring the sensor emission over several tens of microseconds after the excitation light is turned off. The lifetime images acquired through the skin are processed by neural network-based models for misalignment-tolerant inference of glucose levels, accurately revealing normal, low (hypoglycemia) and high (hyperglycemia) concentration ranges. Using a 1-mm thick skin phantom mimicking the optical properties of human skin, we performed in vitro testing of the PLI using glucose-spiked samples, yielding 88.8% inference accuracy, also showing resilience to random and unknown misalignments within a lateral distance of ~4.7 mm with respect to the position of the insertable sensor underneath the skin phantom. Furthermore, the PLI accurately identified larger lateral misalignments beyond 5 mm, prompting user intervention for re-alignment. The misalignment-resilient glucose concentration inference capability of this compact and cost-effective phosphorescence lifetime imager makes it an appealing wearable diagnostics tool for real-time tracking of glucose and other biomarkers.