通过从2D kV图像合成患者特异性3D CT图像实现无额外成像剂量的精准患者摆位
Accurate Patient Alignment without Unnecessary Imaging Dose via Synthesizing Patient-specific 3D CT Images from 2D kV Images
AI总结:
本文提出基于分层ViT的双模型框架,仅用2D kV图像实时合成患者特异性3D CT,在头颈癌患者上实现高图像质量、高剂量学精度和亚毫米级摆位精度,从而在最小化成像剂量的同时提升放射治疗摆位准确性。
摘要:
在放射治疗中,当无法使用3D机载成像(OBI)时,通常采用2D正交投影kV图像进行患者摆位。但由于患者解剖结构被投影到2D平面上,肿瘤可见性受到限制,可能导致较大的摆位误差。在配备3D-OBI(如锥形束CT,CBCT)的治疗室中,CBCT的视野(FOV)有限且成像剂量不必要地偏高,因此不利于儿科患者。解决这一困境的方法是利用治疗位置获取的kV图像重建3D CT。本文提出了一种基于分层ViT模块构建的双模型框架。与概念验证方法不同,该框架将kV图像作为唯一输入,并能在实时(毫秒级)内合成精确的全尺寸3D CT。我们在10例头颈部(H&N)癌症患者上验证了所提方法的可行性,评估指标包括图像质量(MAE:<45HU)、剂量学精度(Gamma通过率(2%/2mm/10%)>97%)以及患者位置不确定性(位移误差:<0.4mm)。所提框架能够生成精确的3D CT,忠实反映实时患者位置,从而显著提高患者摆位精度,保持成像剂量最小化,并维持治疗的真实性。
英文摘要:
In radiotherapy, 2D orthogonally projected kV images are used for patient alignment when 3D-on-board imaging(OBI) unavailable. But tumor visibility is constrained due to the projection of patient's anatomy onto a 2D plane, potentially leading to substantial setup errors. In treatment room with 3D-OBI such as cone beam CT(CBCT), the field of view(FOV) of CBCT is limited with unnecessarily high imaging dose, thus unfavorable for pediatric patients. A solution to this dilemma is to reconstruct 3D CT from kV images obtained at the treatment position. Here, we propose a dual-models framework built with hierarchical ViT blocks. Unlike a proof-of-concept approach, our framework considers kV images as the solo input and can synthesize accurate, full-size 3D CT in real time(within milliseconds). We demonstrate the feasibility of the proposed approach on 10 patients with head and neck (H&N) cancer using image quality(MAE: <45HU), dosimetrical accuracy(Gamma passing rate (2%/2mm/10%)>97%) and patient position uncertainty(shift error: <0.4mm). The proposed framework can generate accurate 3D CT faithfully mirroring real-time patient position, thus significantly improving patient setup accuracy, keeping imaging dose minimum, and maintaining treatment veracity.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。