arXiv 2405.18403physics.ins-det
收藏

SBC-LAr10 闪烁气泡室用 VUV4 批量表征

Batch VUV4 Characterization for the SBC-LAr10 scintillating bubble chamber

H. Hawley-Herrera, E. Alfonso-Pita, E. Behnke, M. Bressler, B. Broerman, K. Clark, J. Corbett, C. E. Dahl, K. Dering, A. de St. Croix, D. Durnford, P. Giampa, J… 展开作者

H. Hawley-Herrera, E. Alfonso-Pita, E. Behnke, M. Bressler, B. Broerman, K. Clark, J. Corbett, C. E. Dahl, K. Dering, A. de St. Croix, D. Durnford, P. Giampa, J. Hall, O. Harris, N. Lamb, M. Laurin, I. Levine, W. H. Lippincott, X. Liu, N. Moss, R. Neilson, M. -C. Piro, D. Pyda, Z. Sheng, G. Sweeney, E. Vázquez-Jáuregui, S. Westerdale, T. J. Whitis, A. Wright, E. Wyman, R. Zhang

AI总结:

该研究对SBC-LAr10闪烁气泡室所用32个VUV4 SiPM进行批量表征,测量关键参数并分析温度依赖性,确保达到实验精度要求,以优化信噪比。

摘要:

闪烁气泡室(SBC)合作组购买了32个滨松 VUV4 硅光电倍增管(SiPM),用于包含10千克液氩的气泡室 SBC-LAr10。采用一种避免使用单光子源的暗计数表征技术,在两个温度下测量了 VUV4 SiPM 的击穿电压($V_{\text{BD}}$)、SiPM 增益($g_{\text{SiPM}}$)、$g_{\text{SiPM}}$ 随电压的变化率($m$)、暗计数率(DCR)以及相关雪崩概率(P$_{\text{CA}}$),并测量了这些参数的温度系数。女王大学开发了一种基于珀尔帖效应的冷却真空室,将四联模块冷却至 $233.15\pm0.2$~K 和 $255.15\pm0.2$~K,平均稳定性为 $\pm20$~mK。开发了一种分析框架,用于将 $V_{\text{BD}}$ 估计到数十毫伏的精度,并将 DCR 估计到接近泊松误差。发现 $V_{\text{BD}}$ 的温度依赖性为 $56\pm2$~mV~K$^{-1}$,所有四联模块的平均 $m$ 为 $(459\pm3(\rm{stat.})\pm23(\rm{sys.}))\times 10^{3}~e^-$~PE$^{-1}$~V$^{-1}$。平均 DCR 温度系数估计为 $0.099\pm0.008$~K$^{-1}$,对应于温度每下降 20~K,DCR 降低 7 倍。P$_{\text{CA}}$ 的平均温度依赖性估计为 $4000\pm1000$~ppm~K$^{-1}$。对于所有其他参数,由所有 SiPM 的平均值估计的 P$_{\text{CA}}$ 比由单个 SiPM 计算的 P$_{\text{CA}}$ 更优,而其他参数则相反。所有估计参数均达到 SBC-LAr10 所需的精度,四联模块将在优化信噪比的条件下使用。

英文摘要:

The Scintillating Bubble Chamber (SBC) collaboration purchased 32 Hamamatsu VUV4 silicon photomultipliers (SiPMs) for use in SBC-LAr10, a bubble chamber containing 10~kg of liquid argon. A dark-count characterization technique, which avoids the use of a single-photon source, was used at two temperatures to measure the VUV4 SiPMs breakdown voltage ($V_{\text{BD}}$), the SiPM gain ($g_{\text{SiPM}}$), the rate of change of $g_{\text{SiPM}}$ with respect to voltage ($m$), the dark count rate (DCR), and the probability of a correlated avalanche (P$_{\text{CA}}$) as well as the temperature coefficients of these parameters. A Peltier-based chilled vacuum chamber was developed at Queen's University to cool down the Quads to $233.15\pm0.2$~K and $255.15\pm0.2$~K with average stability of $\pm20$~mK. An analysis framework was developed to estimate $V_{\text{BD}}$ to tens of mV precision and DCR close to Poissonian error. The temperature dependence of $V_{\text{BD}}$ was found to be $56\pm2$~mV~K$^{-1}$, and $m$ on average across all Quads was found to be $(459\pm3(\rm{stat.})\pm23(\rm{sys.}))\times 10^{3}~e^-$~PE$^{-1}$~V$^{-1}$. The average DCR temperature coefficient was estimated to be $0.099\pm0.008$~K$^{-1}$ corresponding to a reduction factor of 7 for every 20~K drop in temperature. The average temperature dependence of P$_{\text{CA}}$ was estimated to be $4000\pm1000$~ppm~K$^{-1}$. P$_{\text{CA}}$ estimated from the average across all SiPMs is a better estimator than the P$_{\text{CA}}$ calculated from individual SiPMs, for all of the other parameters, the opposite is true. All the estimated parameters were measured to the precision required for SBC-LAr10, and the Quads will be used in conditions to optimize the signal-to-noise ratio.