Self-sensing with hollow cylindrical transducers for histotripsy enhanced aspiration mechanical thrombectomy applications

Li Gong, Alex R. Wright, Kullervo Hynynen, David E. Goertz

AI总结:

本研究提出利用空心圆柱换能器自传感电压信号检测管腔内空化,用于组织摧毁术增强抽吸取栓,实验验证了其高灵敏度可行性。

摘要:

为应对血管内机械取栓装置面临的现有挑战,正在开发一种新型超声增强抽吸方法,利用空化作用机械降解血栓。该方法在毫米级空心圆柱换能器内产生驻波,以生成足以对位于换能器管腔内的血栓实施组织摧毁术并产生显著损伤的高压力。本研究旨在通过分析治疗脉冲期间换能器两端的电压信号,评估自传感空化检测的可行性。换能器管腔的外部超声成像验证了空化检测。管腔内血栓物质的存在也改变了阻抗。在充满水的管腔中改变驱动电压的实验表明,在空化阈值以上,脉冲主体和振铃部分的包络相对幅度发生变化,频谱域也出现变化。特别是,宽带和超谐波信号在空化阈值以上均显示幅度增加。在管腔内处理血栓时,也观察到空化存在时相似的时域和频域电压信号变化。这项工作展示了一种高度灵敏的管腔内空化检测方法,使得无需在导管配置中增加额外传感器即可利用易于获取的信号进行监测。这些发现对提高超声增强抽吸取栓手术的有效性具有重要潜力。

英文摘要:

To address existing challenges with intravascular mechanical thrombectomy devices, a novel ultrasound-enhanced aspiration approach is being developed to mechanically degrade clots using cavitation. This method employs standing waves within a mm-scale hollow cylindrical transducer to generate high pressures sufficient to perform histotripsy on clots situated within the transducer lumen and generate substantial lesions. The objective of this study is to assess the feasibility of self-sensing cavitation detection by analyzing voltage signals across the transducer during treatment pulses. External ultrasound imaging of the transducer lumen validated cavitation detection. Impedance was also altered by the presence of clot material within the lumen. Experiments varying the driving voltage in water-filled lumens demonstrated changes in the relative amplitudes of the envelopes of the pulse body and ringdown portions of the voltage signals above the cavitation threshold, as well as changes in the spectral domain. In particular both broadband and ultraharmonic signals showed an increase in amplitude above the cavitation threshold. Similar temporal and spectral voltage signal changes in the presence of cavitation were also observed when treating clots within the lumen. This work demonstrates a highly sensitive method for detecting cavitation within the lumen, enabling monitoring with readily acquired signals without additional sensors in the catheter configuration. These findings hold significant potential for improving the efficacy of ultrasound-enhanced aspiration thrombectomy procedures.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑