双换能器配置下的焦点体积、声辐射力与应变
Focal Volume, Acoustic Radiation Force, and Strain in Two-Transducer Regimes
AI总结:
本文提出双1 MHz正交换能器方案,将TUS空间分辨率提升近40倍、靶体积降至0.24 mm³,并实现局域化定向声辐射力与选择性正负压力施加,为精细尺度神经调控研究提供新工具。
摘要:
经颅聚焦超声刺激(TUS)在治疗神经系统疾病的无创神经调控方面具有广阔前景。大多数临床相关靶点位于大脑深处(接近或处于其几何中心),周围环绕着需要避免临床干预的其他敏感区域。然而,在TUS中,以提高空间分辨率为目的而增加频率会降低有效穿透深度。我们表明,通过使用一对1 MHz、正交布置的换能器,我们将每个换能器单独所能提供的空间分辨率提高了近40倍,实现了0.24 mm³的亚立方毫米靶体积。我们表明,正交放置的换能器会产生高度局域化的驻波,其声辐射力(ARF)在靶点附近排列成周期性的压缩区和拉伸区。我们进一步展示了正交配置的扩展能力,即能够对靶点施加选择性压力——要么为正压,要么为负压,但不能同时施加两者。最后,我们分享了初步发现:应变既可以由粒子运动产生,也可以由ARF产生,前者在焦点处达到最大值,而后者在焦点处保持为零并在焦点周围达到最大值。随着该领域正在通过阐明超声参数与神经反应之间的映射来研究TUS中的相互作用机制,正交换能器扩展了我们的工具箱,使得能够在更精细的空间分辨率下开展这些研究,并具备局域化且具有方向性(压缩与拉伸)的ARF以及在靶点施加选择性压力的能力。
英文摘要:
Transcranial focused ultrasound stimulation (TUS) holds promise for non-invasive neural modulation in treating neurological disorders. Most clinically relevant targets are deep within the brain (near or at its geometric center), surrounded by other sensitive regions that need to be spared clinical intervention. However, in TUS, increasing frequency with the goal of improving spatial resolution reduces the effective penetration depth. We show that by using a pair of 1 MHz, orthogonally arranged transducers we improve the spatial resolution afforded by each of the transducers individually, by nearly 40 fold, achieving a sub-cubic millimeter target volume of $0.24\ mm^3$. We show that orthogonally placed transducers generate highly localized standing waves with Acoustic Radiation Force (ARF) arranged into periodic regions of compression and tension near the target. We further present an extended capability of the orthogonal setup, which is to impart selective pressures--either positive or negative, but not both--on the target. Lastly, we share our preliminary findings that strain can arise from both particle motion and ARF with the former reaching its maximum value at the focus, and the latter remaining null at the focus and reaching its maximum around the focus. As the field is investigating the mechanism of interaction in TUS by way of elucidating the mapping between ultrasound parameters and neural response, orthogonal transducers expand our toolbox by making it possible to conduct these investigations at much finer spatial resolutions, with localized and directed (compression vs. tension) ARF and the capability of applying selective pressures at the target.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。