考虑非理想梯度性能的MR序列设计
MR sequence design to account for non-ideal gradient performance
AI总结:
针对传统MRI预加重补偿涡流导致硬件规格下调、不适用于低性能设备的问题,提出基于优化的预补偿梯度波形设计方法,结合系统缺陷模型优化波形,在7T系统上实现鬼影伪影平均降低77%,提升低成本硬件成像性能。
摘要:
传统上,MRI系统被设计为可产生接近理想化的性能,这催生了简化的脉冲序列设计理念。其中一个例子是对梯度场产生的涡流的控制:通常这些涡流通过对所需波形施加预加重来补偿。这一过程通常对脉冲序列设计者是不可见的,使得他们可以假定实际获得的梯度波形会符合预期。这种方式虽然便捷,但要求向终端用户公开的系统规格被大幅下调,因为预加重会给波形增加额外的开销。这种策略对于性能较低或资源有限的硬件而言并不可取。\n为此,我们提出一种基于优化的方法来设计预补偿梯度波形,该方法具备两个特点:(i)明确遵循硬件约束;(ii)通过直接校正k空间采样提升成像性能。我们通过纳入系统缺陷模型对梯度波形进行数值优化。研究首先采用指数涡流模型开展仿真研究,随后在7T MRI系统上使用经验梯度系统传递函数进行实验验证。\n我们提出的方法找到了可产生可忽略重建误差同时满足梯度系统限制的解决方案,即便在经典预加重会产生不可行结果的情况下依然有效。实验观察到EPI成像的鬼影伪影大幅减少,其中体模中的鬼影幅度平均降低了77%。这项工作证明了梯度波形的数值优化在给定系统缺陷模型时可大幅提升图像质量。虽然该方法在实现时灵活性有限,但它能够实现更高效的硬件利用,并且可能对于最大化低成本系统的性能尤为重要。
英文摘要:
MRI systems are traditionally engineered to produce close to idealized performance, enabling a simplified pulse sequence design philosophy. An example of this is control of eddy currents produced by gradient fields; usually these are compensated by pre-emphasizing demanded waveforms. This process typically happens invisibly to the pulse sequence designer, allowing them to assume achieved gradient waveforms will be as desired. Whilst convenient, this requires system specifications exposed to the end-user to be substantially down-rated, since pre-emphasis adds an extra overhead to the waveforms. This strategy is undesirable for lower performance or resource-limited hardware. Instead, we propose an optimization-based method to design pre-compensated gradient waveforms that: (i) explicitly respect hardware constraints and (ii) improve imaging performance by correcting k-space samples directly. Gradient waveforms are numerically optimized by including a model for system imperfections. This is investigated in simulation using an exponential eddy current model, then experimentally using an empirical gradient system transfer function on a 7T MRI system. Our proposed method discovers solutions that produce negligible reconstruction errors while satisfying gradient system limits, even when classic pre-emphasis produces infeasible results. Substantial reduction in ghosting artefacts from EPI imaging was observed, including an average reduction of 77% in ghost amplitude in phantoms. This work demonstrates numerical optimization of gradient waveforms, yielding substantially improved image quality when given a model for system imperfections. While the method as implemented has limited flexibility, it could enable more efficient hardware usage, and may prove particularly important for maximizing performance of lower-cost systems.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。