MedAustron 临床碳离子笔形束的微剂量学——第 1 部分:实验表征
Microdosimetry of a clinical carbon-ion pencil beam at MedAustron -- Part 1: experimental characterization
AI总结:
本文利用固态硅微剂量计实验表征 MedAustron 碳离子笔形束的微剂量谱,发现离轴距离影响碎片贡献和辐射质量,对相对生物有效性有重要意义。
摘要:
本文利用微型固态硅微剂量计对 MedAustron 单能碳离子笔形束的微剂量谱进行了表征,以评估不同碎片横向分布对微剂量谱的影响。微剂量计固定在某一深度,然后以约 2 mm 的步长从中心束轴横向移动。测量在水体模中不同深度处沿水平和垂直方向进行。在布拉格峰之后远端剂量下降区的一个位置,使用整个 y 值范围或子范围(假定主要对应初级粒子的贡献)推导了频率平均线能量和剂量平均线能量。测量的微剂量谱在距束流中心轴 4 mm 以内没有显著变化。对于距中心轴超过 4 mm 的横向位置,由于次级碎片的部分剂量增加,谱中较低线能量部分的相对贡献随横向距离增加而增加。对于每个部分贡献,平均值 yF 和 yD 几乎恒定。然而,当所有粒子一起考虑时,yF 和 yD 的平均值随距轴距离变化,因为这两个组分的剂量分数在距轴最远的垂直位置分别变化了 30% 和 10%。在布拉格峰远端部分的下游,观察到微剂量谱中的特征性结构,强烈表明存在氦和硼碎片。我们能够研究辐射质量随离轴位置的变化。这些测量强调了束流内辐射质量的变化,这对相对生物有效性具有影响。
英文摘要:
This paper characterizes the microdosimetric spectra of a single-energy carbon-ion pencil beam at MedAustron using a miniature solid-state silicon microdosimeter to estimate the impact of the lateral distribution of the different fragments on the microdosimetric spectra. The microdosimeter was fixed at one depth and then laterally moved away from the central beam axis in steps of approximately 2 mm. The measurements were taken in both horizontal and vertical direction in a water phantom at different depths. In a position on the distal dose fall-off beyond the Bragg peak, the frequency-mean and the dose-mean lineal energies were derived using either the entire range of y-values, or a sub-range of y values, presumingly corresponding mainly to contributions from primary particles. The measured microdosimetric spectra do not exhibit a significant change up to 4 mm away from the beam central axis. For lateral positions more than 4 mm away from the central axis, the relative contribution of the lower lineal-energy part of the spectrum increases with lateral distance due to the increased partial dose from secondary fragments. The average values yF and yD are almost constant for each partial contribution. However, when all particles are considered together, the average value of yF and yD varies with distance from the axis due to the changing dose fractions of these two components varying by 30 % and 10 % respectively up to the most off axis vertical position. Characteristic features in the microdosimetric spectra providing strong indications of the presence of helium and boron fragments have been observed downstream of the distal part of the Bragg peak. We were able to investigate the radiation quality as function of off-axis position. These measurements emphasize variation of the radiation quality within the beam and this has implications in terms of relative biological effectiveness.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。