采用图像处理技术检测听力正常婴儿的听性脑干反应波峰
Detection of Auditory Brainstem Response Peaks Using Image Processing Techniques in Infants with Normal Hearing Sensitivity
AI总结:
本研究提出一种基于图像处理技术的听性脑干反应(ABR)波峰自动检测方法,在26名听力正常婴幼儿的ABR数据上验证了有效性,可提升听力学测试的准确性与效率。
摘要:
引言:听性脑干反应(ABR)被用于检测听力正常儿童的脑干水平外周听觉神经系统完整性。听觉诱发电位(AEP)由声学刺激产生。解读这些波形需要专业能力,以避免对听力问题的误诊。借助计算机视觉实现ABR测试标记的自动化,或可减少人为误差。方法:本研究采用26名1至20个月大、双耳听力正常儿童的ABR测试结果,提出了一种自动计算不同强度(单位:分贝)波形波峰的新方法。该流程包括:使用Color Thresholder(颜色阈值分割器)方法从Audera设备获取波形图像,利用Image Region Analyzer(图像区域分析器)应用将每个波形分割为单波形图像,通过图像处理(IP)技术将所有波形图像转换为波形,最后计算每个波形的波峰潜伏期,供听力学家用于疾病诊断。结果:图像处理技术在诊断领域检测1波、3波和5波的准确率分别为0.82、0.98和0.98,对1波、3波和5波的精确率分别为0.32、0.97和0.87。该评估在阈值检测部分也表现良好,ABR波形的正确检测率达82.7%。结论:我们的研究结果表明,通过使用技术自动检测和标记ABR波形,可以让听力学测试组合套件更准确、更快速且更少出错。
英文摘要:
Introduction: The auditory brainstem response (ABR) is measured to find the brainstem-level peripheral auditory nerve system integrity in children having normal hearing. The Auditory Evoked Potential (AEP) is generated using acoustic stimuli. Interpreting these waves requires competence to avoid misdiagnosing hearing problems. Automating ABR test labeling with computer vision may reduce human error. Method: The ABR test results of 26 children aged 1 to 20 months with normal hearing in both ears were used. A new approach is suggested for automatically calculating the peaks of waves of different intensities (in decibels). The procedure entails acquiring wave images from an Audera device using the Color Thresholder method, segmenting each wave as a single wave image using the Image Region Analyzer application, converting all wave images into waves using Image Processing (IP) techniques, and finally calculating the latency of the peaks for each wave to be used by an audiologist for diagnosing the disease. Findings: Image processing techniques were able to detect 1, 3, and 5 waves in the diagnosis field with accuracy (0.82), (0.98), and (0.98), respectively, and its precision for waves 1, 3, and 5, were respectively (0.32), (0.97) and (0.87). This evaluation also worked well in the thresholding part and 82.7 % correctly detected the ABR waves. Conclusion: Our findings indicate that the audiology test battery suite can be made more accurate, quick, and error-free by using technology to automatically detect and label ABR waves.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。