超声弹性成像用于预防乳房植入物破裂:检测硬度随植入时间增加的现象

Ultrasonic elastography for the prevention of breast implant rupture: Detection of an increase with stiffness over implantation time

Laetitia Ruffenach, Dimitri Heintz, Claire Villette, Charlène Cosentino, Denis Funfschilling, Frédéric Bodin, Nadia Bahlouli, Simon Chatelin

AI总结:

本研究通过ARFI超声弹性成像及MSI、LC-HRMS分析35个乳房取出物,发现植入8年后硬度和剪切模量显著升高,并揭示胆固醇代谢物降解凝胶,从而为定量评估破裂风险和更换时机提供依据。

摘要:

乳房植入物在乳腺癌切除术后被广泛使用,并且必须定期更换以避免破裂。迄今为止,尚没有定量标准来辅助这一决策。植入物凝胶和膜的力学演变仍未得到充分研究,尽管这种演变可能导致早期破裂。在本研究中,对在患者体内植入时间最长达17年的35个乳房取出物进行了表征,方法是对其力学性能进行离体测量。利用声辐射力脉冲(ARFI)超声弹性成像这一用于无损力学表征的成像方法,研究观察到取出物的硬度有所增加。这种增加与植入持续时间相关,主要发生在植入8年之后。由于剪切模量最高增加至近3倍,植入物柔韧性的丧失很可能导致其破裂风险显著增加。通过质谱成像(MSI)以及液相色谱联用高分辨质谱(LC-HRMS)对取出物中的凝胶进行补充分析,证实存在源自乳房组织的胆固醇代谢物;这些代谢物极有可能穿过植入物的膜,并很可能使凝胶降解。通过观察患者体内正在发挥作用的物理化学机制所造成的后果,本研究表明,超声弹性成像可在体内用作乳房植入物破裂风险的定量指标,并帮助诊断是否需要对其进行更换。

英文摘要:

Breast implants are widely used after breast cancer resection and must be changed regularly to avoid a rupture. To date, there are no quantitative criteria to help this decision. The mechanical evolution of the gels and membranes of the implants is still under investigated, although it can lead to early rupture. In this study, 35 breast explants having been implanted in patients for up to 17 years were characterized by ex vivo measurements of their mechanical properties. Using Acoustic Radiation Force Impulse (ARFI) ultrasound elastography, an imaging method for non-destructive mechanical characterization, an increase in the stiffness of the explants has been observed. This increase was correlated with the implantation duration, primarily after 8 years of implantation. With an increase of the shear modulus of up to a factor of nearly 3, the loss of flexibility of the implants is likely to lead to a significant increase of their risk of rupture. A complementary analysis of the gel from the explants by mass spectrometry imaging (MSI) and liquid chromatography coupled to high resolution mass spectrometry (LC-HRMS) confirms the presence of metabolites of cholesterol originating from the breast tissues, which most likely crossed the membrane of the implants and most likely degrades the gel. By observing the consequences of the physical-chemical mechanisms at work within patients, this study shows that ultrasound elastography could be used in vivo as a quantitative indicator of the risk of breast implant rupture and help diagnose their replacement.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑