多模式目标透镜用于动量显微镜和XPEEM:实验
Multimode Objective Lens for Momentum Microscopy and XPEEM: Experiments
AI总结:
本文提出了一种多模式目标透镜,用于XPEEM和动量显微镜,通过实验验证其在不同光能范围内的性能,提升了成像分辨率并减少了空间电荷效应。
摘要:
一种新的目标透镜最近被提出用于X射线光电子能谱显微镜(XPEEM)和动量显微镜中。在提取器内添加一个与提取器同心的环形电极,可以对样品与提取器之间的场进行整形。在该间隙中形成一个透镜场,可将作用于样品的场强度降低至最大一个数量级。这减轻了场发射的风险,特别是对于具有锐利边缘的切割样品。一个延缓场可以将所有慢电子重新定向,从而消除空间电荷相互作用的主要贡献。在这里,我们介绍了新透镜的第一个实验研究,考察其在由高阶谐波生成(HHG)源产生的极紫外光到软X射线和硬X射线两种同步辐射设施中的性能。在无电极的区域中,间隙透镜能够实现高达23毫米的工作距离。减少的像差允许在k空间和实空间成像中获得更大的视野,其分辨率与传统阴极透镜相当。然而,场强则小一个数量级。零场模式使研究三维结构的物体成为可能,因此对小切割样品以及涉及顶部电极的原位设备都是有益的。排斥模式减少了空间电荷效应,但导致k场直径减小。这种减小范围从在硬X射线能量下为10%到在XUV范围内为50%。可用电能区间也减少了一半。在飞行时间XPEEM模式中,原始数据显示出250纳米的分辨率,通过数据处理可提高到优于100纳米。
英文摘要:
A new type of objective lens has recently been proposed for use in X-ray photoemission electron microscopes (XPEEMs) and momentum microscopes. Adding a ring electrode concentric with the extractor allows the field in the gap between the sample and the extractor to be shaped. Forming a lens field in this gap reduces the field strength at the sample by up to an order of magnitude. This mitigates the risk of field emission, particularly for cleaved samples with sharp edges. A retarding field can redirect all slow electrons, thus eliminating the primary contribution to the space-charge interaction. Here we present the first experimental investigation of the new lens, examining its performance at photon energies ranging from the extreme ultraviolet produced by a high-harmonic generation (HHG)-based source to soft and hard X-rays at two synchrotron facilities. The gap lens in a region without electrodes enables large working distances up to 23 mm. Reduced aberrations allow for larger fields of view in both k-space and real-space imaging, with resolutions comparable to those of conventional cathode lenses. However, field strengths are an order of magnitude smaller. The zero-field mode enables the study of 3D structured objects and is therefore beneficial for small cleaved samples as well as for operando devices involving top electrodes. The repeller mode reduces space-charge effects, but results in a smaller k-field diameter. This reduction ranges from 10% at hard X-ray energies to 50% in the XUV range. The usable energy interval is also reduced by a factor of two. In time-of-flight XPEEM mode the raw data show a resolution of 250 nm, which can be improved to better than 100 nm through data processing.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。