利用机械原位加载台处理低对比度聚合物样品nanoCT重建中的缺失角区

Dealing with missing angular sections in nanoCT reconstructions of low contrast polymeric samples employing a mechanical in situ loading stage

Rafaela Debastiani, Chantal Miriam Kurpiers, Enrico Domenico Lemma, Ben Breitung, Martin Bastmeyer, Ruth Schwaiger, Peter Gumbsch

AI总结:

针对Xradia 810 Ultra原位加载导致nanoCT仅能采集140°、低对比度聚合物重建失真的问题,本文通过样品定位与旋转策略,将缺失细节限制在约20°楔形区并实现准确重建。

摘要:

尽管原位实验在超材料或具有复杂微观结构材料的(力学)评估中日益重要,但此类实验中的成像条件往往具有挑战性。实验室计算机断层扫描系统Xradia 810 Ultra支持对样品进行原位(延时)力学测试。然而,该系统的原位加载装置将图像采集角度限制在140°。对于低对比度聚合物材料,这一有限的采集角度会导致信息增益较低的区域,从而阻碍使用滤波反投影算法准确重建数据。在此,我们展示了如何通过选择合适的样品位置来提高信息增益。一个低对比度聚合物四面体微晶格样品和一个具有特定结构的样品均在140°和180°下进行扫描,结果表明140°重建中缺失的结构细节局限于约20°的角度楔形区。根据样品几何形状和结构,在原位实验中采用简单策略即可准确重建数据。对于四面体微晶格,仅将样品简单旋转90°即可提供足够的X射线吸收,从而实现几何形状的准确重建。

英文摘要:

While in situ experiments are gaining importance for the (mechanical) assessment of metamaterials or materials with complex microstructures, imaging conditions in such experiments are often challenging. The lab-based computed tomography system Xradia 810 Ultra allows for the in situ (time lapsed) mechanical testing of samples. However, the in situ loading setup from this system limits the image acquisition angle to 140$^\circ$. For low contrast polymeric materials, this limited acquisition angle leads to regions of low information gain, thus preventing an accurate reconstruction of the data using a filtered back projection algorithm. Here we demonstrate how the information gain can be improved by selecting an appropriate position of the sample. A low contrast polymeric tetrahedral microlattice sample and a specifically structured sample, both scanned over 140$^\circ$ and 180$^\circ$, demonstrate that the missing structural details in the 140$^\circ$ reconstruction are limited to an angular wedge of about 20$^\circ$. Depending on the sample geometry and structure, applying simple strategies for the in situ experiments allows accurate reconstruction of the data. For the tetrahedral microlattice, a simple rotation of the sample by 90$^\circ$ provides enough X-ray absorption for an accurate reconstruction of the geometry.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑