使用宽带光源进行望远镜透射率测量的准直光束投影仪(CBP)的设计与性能
Design and performance of a Collimated Beam Projector for telescope transmission measurement using a broadband light source
AI总结:
本文介绍便携式准直光束投影仪原型,采用宽带氙灯与单色仪测量望远镜波长吞吐量,在ZTF的g、r波段分别实现1.2%和2.3%精度。
摘要:
Ia型超新星是研究近邻宇宙中暗能量最直接的宇宙学探针,而天文仪器的测光校准仍是其系统不确定性的主要来源之一。为解决这一问题,近期进展引入了准直光束投影仪(Collimated Beam Projectors,CBP),旨在通过精确测量望远镜吞吐量随波长的变化来增强校准。本工作描述了一台便携式CBP原型机的性能。实验装置采用宽带氙灯光源,替代了更常用但要求高得多的高功率激光源;该光源与单色仪耦合,单色仪在积分球内发射光,并用光电二极管和光谱仪进行监测。光被注入CBP望远镜的焦点,由望远镜将准直光束投射到太阳能电池上,该太阳能电池的量子效率已通过与NIST校准的光电二极管比较获得。通过比较CBP光电二极管与太阳能电池中的光电流信号,计算了所实现的吞吐量和信噪比。我们证明,在当前开发状态下,该原型机能够在合理的积分时间内,对ZTF测光滤光片系统的积分g波段和r波段分别达到1.2%和2.3%的精度。g波段和r波段的中心波长测定精度分别保持在约0.91 nm和约0.58 nm以下。由滤光片吞吐量测量引起的预期测光不确定性在零点星等上约为5 mmag。文中还讨论了若干可直接用于升级当前装置的改进途径。
英文摘要:
Type Ia supernovae are the most direct cosmological probe to study dark energy in the recent Universe, for which the photometric calibration of astronomical instruments remains one major source of systematic uncertainties. To address this, recent advancements introduce Collimated Beam Projectors (CBP), aiming to enhance calibration by precisely measuring a telescope's throughput as a function of wavelength. This work describes the performance of a prototype portable CBP. The experimental setup consists of a broadband Xenon light source replacing a more customary but much more demanding high-power laser source, coupled with a monochromator emitting light inside an integrating sphere monitored with a photodiode and a spectrograph. Light is injected at the focus of the CBP telescope projecting a collimated beam onto a solar cell whose quantum efficiency has been obtained by comparison with a NIST-calibrated photodiode. The throughput and signal-to-noise ratio achieved by comparing the photocurrent signal in the CBP photodiode to the one in the solar cell are computed. We prove that the prototype, in its current state of development, is capable of achieving 1.2 per cent and 2.3 per cent precision on the integrated g and r bands of the ZTF photometric filter system respectively, in a reasonable amount of integration time. Central wavelength determination accuracy is kept below $\sim$ {0.91} nm and $\sim$ {0.58} nm for g and r bands. The expected photometric uncertainty caused by filter throughput measurement is approximately 5 mmag on the zero-point magnitude. Several straightforward improvement paths are discussed to upgrade the current setup.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。