使用期望效用(EU)评估用于乳腺X线筛查的人工智能排除设备
Use of Expected Utility (EU) to Evaluate Artificial Intelligence-Enabled Rule-Out Devices for Mammography Screening
AI总结:
本文提出用PPV/NPV和期望效用(EU)替代敏感度与特异度来评估AI乳腺X线筛查排除设备,并通过美欧两项回顾性研究验证EU能更准确反映筛查性能变化。
摘要:
背景:人工智能(AI)驱动的排除设备可以自主地将不太可能患有癌症的患者影像从放射科医生的审阅中移除。许多已发表的研究通过将AI回顾性地应用于大型数据集来评估这类设备,并使用敏感度和特异度作为性能指标。然而,这些指标存在根本性缺陷,因为在AI排除应用中,它们必然会发生相反方向的变化。方法:我们回顾了两种用于比较“放射科医生使用排除设备”与“放射科医生不使用设备”工作流程筛查性能的指标:阳性/阴性预测值(PPV/NPV)和期望效用(EU)。我们将这两种方法应用于一项近期研究,该研究使用回顾性美国数据集报告了放射科医生使用设备工作流程的性能改善。随后,我们将EU方法应用于一项欧洲研究,基于不同AI阈值下报告的召回率和癌症检出率,比较不同阈值之间的潜在效用。结果:对于美国研究,在所报告的任何算法阈值下,PPV/NPV和EU均无法证明存在显著改善。对于使用欧洲数据的研究,我们发现随着AI排除更多患者(包括假阴性病例),EU降低,并降低了整体筛查性能。结论:由于回顾性模拟研究设计的性质,敏感度和特异度在评估排除设备时可能存在歧义。我们表明,使用PPV/NPV或EU可以消除这种歧义。EU方法仅需召回率和癌症检出率即可应用,这在筛查性乳腺X线检查中非召回患者通常缺乏真实标签的情况下十分方便。
英文摘要:
Background: An artificial intelligence (AI)-enabled rule-out device may autonomously remove patient images unlikely to have cancer from radiologist review. Many published studies evaluate this type of device by retrospectively applying the AI to large datasets and use sensitivity and specificity as the performance metrics. However, these metrics have fundamental shortcomings because they are bound to have opposite changes with the rule-out application of AI. Method: We reviewed two performance metrics to compare the screening performance between the radiologist-with-rule-out-device and radiologist-without-device workflows: positive/negative predictive values (PPV/NPV) and expected utility (EU). We applied both methods to a recent study that reported improved performance in the radiologist-with-device workflow using a retrospective U.S. dataset. We then applied the EU method to a European study based on the reported recall and cancer detection rates at different AI thresholds to compare the potential utility among different thresholds. Results: For the U.S. study, neither PPV/NPV nor EU can demonstrate significant improvement for any of the algorithm thresholds reported. For the study using European data, we found that EU is lower as AI rules out more patients including false-negative cases and reduces the overall screening performance. Conclusions: Due to the nature of the retrospective simulated study design, sensitivity and specificity can be ambiguous in evaluating a rule-out device. We showed that using PPV/NPV or EU can resolve the ambiguity. The EU method can be applied with only recall rates and cancer detection rates, which is convenient as ground truth is often unavailable for non-recalled patients in screening mammography.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。