具有系统内校准的测量驱动晶体间散射恢复算法的开发与验证

Development and validation of a measurement-driven inter crystal scatter recovery algorithm with in-system calibration

Katrin Herweg, Volkmar Schulz, David Schug

AI总结:

本文提出一种结合系统内测量驱动校准与光模式拟合的PET晶体间散射恢复算法,并在Geant4模拟和一对一耦合探测器上验证其晶体识别与能量恢复性能。

摘要:

在PET中,被探测到的伽马光子有很高比例会在闪烁体内发生康普顿散射。闪烁体模块通常由光学隔离的晶体制成。根据入射角和闪烁体几何结构的不同,这可能导致晶体间散射(ICS)事件,即能量沉积在探测器中的两个或更多晶体内,而常见的定位和重建算法无法解析这类事件。因此,ICS事件会降低重建图像的空间分辨率和信噪比。我们希望通过从ICS事件中恢复各个晶体及其相应的能量沉积来应对这一挑战。这些信息最终可输入图像重建框架。在本研究中,我们基于一种将读出通道与每个晶体耦合(一对一耦合)的探测器建立了算法,该算法结合测量驱动校准和拟合程序,以根据测得的光模式实现晶体相互作用的恢复。我们使用Geant4模拟,通过将恢复事件与模拟真值进行比较,验证并优化了该方法。结果表明,在性能最佳的算法版本下,95%至97%的模拟事件能够识别所有正确晶体,并且晶体能量以及事件能量总和均可得到充分恢复。对于事件能量总和,96%的事件可实现小于5%的偏差。总体而言,所开发的ICS恢复算法已在一对一耦合探测器上得到成功验证。未来应可应用于其他探测器配置,并将对此展开研究。此外,正在研究利用新的晶体相互作用信息来确定最可能的首次相互作用晶体,以提高PET重建的效率和信噪比。

英文摘要:

In PET a high percentage of gamma photons being detected undergo Compton scattering in the scintillator. Scintillator blocks are often built from optically isolated crystals. Depending on the angle of incidence and the scintillator geometry this might lead to inter crystal scatter (ICS) events, where energy is deposited in two or more crystals in the detector, which common positioning and reconstruction algorithms cannot resolve. Therefore, ICS events worsen the spatial resolution and the signal-to-noise ratio in the reconstructed image. We want to address this challenge by recovering individual crystals from ICS events with their corresponding energy deposits. This information could ultimately be fed into an image reconstruction framework. In this work, we established an algorithm based on a detector that couples a readout channel to each crystal (one-to-one coupling), which combines a measurement-driven calibration and a fitting routine to achieve the recovery of crystal interactions from measured light patterns. Using Geant4 simulations, we validated and optimized this approach by comparing the recovered events to the simulation ground truth. We showed that, with the best performing algorithm versions, all correct crystals could be identified for 95-97% of the simulated events and the crystal energies as well as the event energy sum could be recovered adequately. For the event energy sum a deviation of less than 5% could be achieved for 96% of all events. Overall, the developed ICS recovery algorithm was successfully validated for one-to-one coupled detector. Future application for other detector configurations should be possible and will be investigated. Additionally, using the new crystal interaction information to determine the most likely first interaction crystal is being examined to improve efficiency and signal-to-noise ratio in the PET reconstruction.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑