利用原子吸收分光光度计测量钆浓度
Gadolinium concentration measurement with an atomic absorption spectrophotometer
AI总结:
本研究利用原子吸收分光光度计,提出掺杂钾、镧及采用钽、钨平台三种方法,以实现水中硫酸钆浓度的准确测量,服务于水切伦科夫探测器的中子标记需求。
摘要:
由于钆(Gd)具有最高的热中子俘获截面,俘获中子后会产生8 MeV的伽马级联,因此有人提议将其溶解于水切伦科夫探测器中,以实现高效的中子标记能力。虽然金属钆不溶于水,但多种钆化合物极易溶解。硫酸钆Gd$_2$(SO$_4$)$_3$已被充分测试,并被提议为最佳候选材料。准确测量其浓度,且不受水中杂质干扰,至关重要。原子吸收分光光度计(AAS)是一种适合此目的的仪器,被广泛用于测量多种元素的浓度。在本研究中,我们描述了使用AAS测量水中硫酸钆浓度的三种不同方法:向样品中掺杂钾和镧,以及采用钽和钨平台。
英文摘要:
Because gadolinium (Gd) has the highest thermal neutron capture cross section, resulting in an 8 MeV gamma cascade upon capture, it has been proposed for dissolution in water Cherenkov detectors to achieve efficient neutron tagging capabilities. While metallic Gd is insoluble in water, several compounds are very easy to dissolve. Gadolinium sulfate, Gd$_2$(SO$_4$)$_3$, has been thoroughly tested and proposed as the best candidate. Accurate measurement of its concentration, free of doubt from impurities in water, is crucial. An atomic absorption spectrophotometer (AAS) is a device that suits this purpose and is widely used to measure the concentration of many elements. In this study, we describe three different approaches to measure Gd sulfate concentrations in water using an AAS: doping samples with potassium and lanthanum, and employing tantalum and tungsten platforms.