皮下悬吊缝合线与人面部软组织生物力学相互作用的有限元分析:一项尸体研究

Finite element analysis of biomechanical interactions of a subcutaneous suspension suture and human face soft-tissue: a cadaver study

Seyed Ali Mousavi, Mohammad Ali Nazari, Pascal Perrier, Masoud Shariat Panahi, John Meadows, Marie-Odile Christen, Ali Mojallal, Yohan Payan

AI总结:

本研究通过有限元模拟分析Silhouette Soft皮下悬吊缝合线与人面部软组织的局部生物力学相互作用,发现增加锥体数量可降低组织应变和应力,模拟位移与实验一致。

摘要:

为了研究面部软组织与Silhouette Soft悬吊缝合线(一种带有CE标志的医疗器械,设计用于面部和颈部软组织的复位)之间的局部相互作用,我们进行了有限元模拟,其中将缝合线模型嵌入到一个三层有限元结构中,该结构考虑了人面部软组织的局部力学组织。在ANSYS中设计了一个局部相互作用的二维轴对称模型,其中组织的几何形状、边界条件和施加的载荷被考虑为局部模拟人面部软组织在面部组织复位中受缝合线约束的情况。Silhouette Soft缝合线由打结的线和锚定在组织中的滑动锥体组成。因此,模拟这些相互作用需要特别注意接触力学的精确建模。由于组织被建模为超弹性材料,面部软组织的位移随缝合线引起的应力强度和锥体数量呈非线性变化。我们的模拟表明,对于4锥缝合线,在2.0N外部载荷下位移为4.35mm,在4.0N下为7.6mm。增加锥体数量导致施加于组织的等效局部应变(约20%)和应力(约60%)降低。模拟的位移与实验观察总体一致。

英文摘要:

In order to study the local interactions between facial soft-tissues and a Silhouette Soft suspension suture, a CE marked medical device designed for the repositioning of soft tissues in the face and the neck, Finite element simulations were run, in which a model of the suture was embedded in a three-layer Finite Element structure that accounts for the local mechanical organization of human facial soft tissues. A 2D axisymmetric model of the local interactions was designed in ANSYS, in which the geometry of the tissue, the boundary conditions and the applied loadings were considered to locally mimic those of human face soft tissue constrained by the suture in facial tissue repositioning. The Silhouette Soft suture is composed of a knotted thread and sliding cones that are anchored in the tissue. Hence, simulating these interactions requires special attention for an accurate modelling of contact mechanics. As tissue is modelled as a hyper-elastic material, the displacement of the facial soft tissue changes in a nonlinear way with the intensity of stress induced by the suture and the number of the cones. Our simulations show that for a 4-cone suture a displacement of 4.35mm for a 2.0N external loading and of 7.6mm for 4.0N. Increasing the number of cones led to the decrease in the equivalent local strain (around 20%) and stress (around 60%) applied to the tissue. The simulated displacements are in general agreement with experimental observations.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑