基于光流的血管呼吸运动补偿
Optical flow-based vascular respiratory motion compensation
AI总结:
本文提出运动相关补偿(MRC)算法,利用血管与可见组织的相关性,结合光流和异常值滤波,实现实时(0.032秒)且准确(误差1.086毫米)的血管呼吸运动补偿,助力机器人辅助介入手术。
摘要:
本文开发了一种新的血管呼吸运动补偿算法,即运动相关补偿(MRC),通过推断不可见血管与可见非血管之间的相关性来进行血管呼吸运动补偿。机器人辅助血管介入手术可以显著减少外科医生的辐射暴露。在机器人辅助图像引导介入中,血管由于呼吸而不断移动/变形,并且除非注射造影剂,否则它们在X射线图像中是不可见的。血管呼吸运动补偿技术预测实时X射线图像中的二维血管路线图。当注射造影剂后血管可见时,基于稀疏的Lucas-Kanade特征跟踪器进行血管呼吸运动补偿。训练一个MRC模型来学习血管和非血管运动之间的相关性。在介入过程中,使用可见组织和训练好的MRC模型来预测不可见的血管。此外,采用基于高斯分布的异常值滤波器进行细化。对体内数据集的实验表明,所提出的方法可以在0.032秒内实现血管呼吸运动补偿,平均误差为1.086毫米。我们实时且准确的血管呼吸运动补偿方法有助于现代血管介入和手术机器人。
英文摘要:
This paper develops a new vascular respiratory motion compensation algorithm, Motion-Related Compensation (MRC), to conduct vascular respiratory motion compensation by extrapolating the correlation between invisible vascular and visible non-vascular. Robot-assisted vascular intervention can significantly reduce the radiation exposure of surgeons. In robot-assisted image-guided intervention, blood vessels are constantly moving/deforming due to respiration, and they are invisible in the X-ray images unless contrast agents are injected. The vascular respiratory motion compensation technique predicts 2D vascular roadmaps in live X-ray images. When blood vessels are visible after contrast agents injection, vascular respiratory motion compensation is conducted based on the sparse Lucas-Kanade feature tracker. An MRC model is trained to learn the correlation between vascular and non-vascular motions. During the intervention, the invisible blood vessels are predicted with visible tissues and the trained MRC model. Moreover, a Gaussian-based outlier filter is adopted for refinement. Experiments on in-vivo data sets show that the proposed method can yield vascular respiratory motion compensation in 0.032 sec, with an average error 1.086 mm. Our real-time and accurate vascular respiratory motion compensation approach contributes to modern vascular intervention and surgical robots.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。