BioGAP:一款面向可穿戴生物信号处理的10核FP能力超低功耗物联网处理器,具备医疗级AFE和BLE连接
BioGAP: a 10-Core FP-capable Ultra-Low Power IoT Processor, with Medical-Grade AFE and BLE Connectivity for Wearable Biosignal Processing
AI总结:
BioGAP平台基于十核超低功耗GAP9 SoC,集成医疗级AFE和BLE,以18.2 mW功耗在SSVEP脑机接口任务中实现3.6 uJ/样本流式处理和15小时续航,验证了边缘近传感器多精度ML处理的可行性。
摘要:
可穿戴生物信号处理应用正在推动面向临床和消费应用的微型化、高能效物联网解决方案取得显著进展。然而,只有通过高能效的边缘处理在传感器附近执行数据处理和机器学习(ML),才能实现向高密度多通道前端的扩展。为应对这些挑战,我们提出了BioGAP,这是一个新颖、紧凑、模块化且轻量(6g)的医疗级生物信号采集与处理平台,由GAP9驱动。GAP9是一款十核超低功耗片上系统(SoC),专为高效的多精度(从FP到激进量化的整数)处理而设计,以满足先进ML和DSP的需求。BioGAP的外形尺寸为16x21x14 mm³,由两块堆叠的PCB组成:一块底板集成了GAP9 SoC、一个具备无线蓝牙低功耗(BLE)能力的SoC、一个电源管理电路和一个加速度计;以及一块包含用于ExG采集的模拟前端(AFE)的屏蔽板。最后,该系统还包括一个可灵活放置的光电容积脉搏波(PPG)PCB,尺寸为9x7x3 mm³,以及一块可充电电池(ϕ 12x5 mm²)。我们在基于稳态视觉诱发电位(SSVEP)的脑机接口(BCI)应用上演示了BioGAP。得益于FFT计算任务上16.7 Mflops/s/mW的效率以及97%的无线带宽缩减,我们在流式模式下实现了3.6 uJ/样本,在板载处理模式下实现了2.2 uJ/样本,总功耗预算仅为18.2 mW,可实现15小时的运行时间。
英文摘要:
Wearable biosignal processing applications are driving significant progress toward miniaturized, energy-efficient Internet-of-Things solutions for both clinical and consumer applications. However, scaling toward high-density multi-channel front-ends is only feasible by performing data processing and machine Learning (ML) near-sensor through energy-efficient edge processing. To tackle these challenges, we introduce BioGAP, a novel, compact, modular, and lightweight (6g) medical-grade biosignal acquisition and processing platform powered by GAP9, a ten-core ultra-low-power SoC designed for efficient multi-precision (from FP to aggressively quantized integer) processing, as required for advanced ML and DSP. BioGAPs form factor is 16x21x14 mm$^3$ and comprises two stacked PCBs: a baseboard integrating the GAP9 SoC, a wireless Bluetooth Low Energy (BLE) capable SoC, a power management circuit, and an accelerometer; and a shield including an analog front-end (AFE) for ExG acquisition. Finally, the system also includes a flexibly placeable photoplethysmogram (PPG) PCB with a size of 9x7x3 mm$^3$ and a rechargeable battery ($ϕ$ 12x5 mm$^2$). We demonstrate BioGAP on a Steady State Visually Evoked Potential (SSVEP)-based Brain-Computer Interface (BCI) application. We achieve 3.6 uJ/sample in streaming and 2.2 uJ/sample in onboard processing mode, thanks to an efficiency on the FFT computation task of 16.7 Mflops/s/mW with wireless bandwidth reduction of 97%, within a power budget of just 18.2 mW allowing for an operation time of 15 h.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。