arXivDaily arXiv每日学术速递 周一至周五更新
arXiv 2307.00099astro-ph.IM
收藏

REACH全球21厘米信号实验的接收机设计

Receiver design for the REACH global 21-cm signal experiment

Ian L. V. Roque, Nima Razavi-Ghods, Steven H. Carey, John A. Ely, Will Handley, Alessio Magro, Riccardo Chiello, Tian Huang, P. Alexander, D. Anstey, G. Bernard… 展开作者

Ian L. V. Roque, Nima Razavi-Ghods, Steven H. Carey, John A. Ely, Will Handley, Alessio Magro, Riccardo Chiello, Tian Huang, P. Alexander, D. Anstey, G. Bernardi, H. T. J. Bevins, J. Cavillot, W. Croukamp, J. Cumner, E. de Lera Acedo, D. I. L. de Villiers, A. Fialkov, T. Gessey-Jones, Q. Gueuning, A. T. Josaitis, G. Kulkarni, S. A. K. Leeney, R. Maiolino, P. D. Meerburg, S. Mittal, M. Pagano, S. Pegwal, C. Pieterse, J. R. Pritchard, A. Saxena, K. H. Scheutwinkel, P. Scott, E. Shen, P. H. Sims, O. Smirnov, M. Spinelli, K. Zarb-Adami

AI总结:

本文介绍REACH全球21厘米信号实验的辐射测量系统设计,通过过约束最小二乘法实现80 mK的校准精度,记录了运往南非前的系统状态。

摘要:

我们详细介绍了旨在实现21厘米中性氢谱线测量的REACH辐射测量系统,内容包括辐射计架构、端到端系统仿真,以及对高可校准系统开发固有挑战的讨论。随后,我们分享了利用过约束最小二乘法计算噪声波参数得到的实验室结果:在与现场使用的天线阻抗相当的定制源上进行5小时积分,校准均方根误差(RMSE)达80 mK。本文记录了2022年12月运往南非前,剑桥地区校准器与数据分析的进展状态。

英文摘要:

We detail the the REACH radiometric system designed to enable measurements of the 21-cm neutral hydrogen line. Included is the radiometer architecture and end-to-end system simulations as well as a discussion of the challenges intrinsic to highly-calibratable system development. Following this, we share laboratory results based on the calculation of noise wave parameters utilising an over-constrained least squares approach demonstrating a calibration RMSE of 80 mK for five hours of integration on a custom-made source with comparable impedance to that of the antenna used in the field. This paper therefore documents the state of the calibrator and data analysis in December 2022 in Cambridge before shipping to South Africa.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑